CLASS II ONGOING Device recall · Reported 3 Apr 2024 · FDA Enforcement Report

Patient Information Center iX (PIC iX) recalled by Philips North America

Philips North America is recalling Patient Information Center iX (PIC iX), distributed nationwide in the US. The recall was initiated on 28 Feb 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: (01)00884838104594(10)4.0.1 (01)00884838104594(10)4.0.2 (01)00884838112070(10)4.1.0 (01)00884838112070(10)4.1.1 (01)00884838121782(10)4.2.0 (01)00884838121782(10)4.2.1; Serial Numbers: 7G6G-7PGV-W 5971-2NL7-Z 6D62-19MD-9 463N-67ME-X 5K2A-6DHP-K 5E42-35J4-P 4E6Z-70HU-B 1T12-0MNA-C 2H34-3MMP-U 6P1E-7HJU-W 3R6X-3XNY-6 7F50-53NK-L 3Z3B-1AHM-H 0C5Y-7FHX-V 0X7R-2AGC-Y 0P5D-5VMH-V 6627-6MKH-U 5J7B-1XPT-T 7U4C-44NB-E 2Y7R-5KGZ-W 1P22-0XNW-P 1004-6RJX-E 1U0P-0FKE-C 396T-5ZH1-5 610R-7AHH-2 0B5K-10GT-X 0Z7K-0VKT-0 2A5D-1NH8-Y 2Y5M-5AGH-E 3153-2END-M 4H2B-5YJP-0 5F6T-6XMZ-Y 5Y3Z-2RHR-N 6B6G-3DKJ-4 7Z05-0UH7-V 5H6A-2ZKV-1 1J61-2KH5-M 612A-2WG5-W 737K-7LPC-G 7Z3Y-0LJ3-2 3T0R-7AHH-X 7A4Y-78J7-V 2P5H-0JLM-Y 002A-5MGJ-6 0X63-19MD-G 221G-5NMA-3 4W6U-0PLF-2 7A3M-6VGP-H 7J0C-69LP-H 0V4G-0EG2-B 2K1C-42GL-Y 4M77-1HM0-G 4R5J-5AGH-B 4T62-19MD-7 4W08-0UH7-U 5Y2X-65P4-E 6H5T-3FHA-4 6K59-0HN3-B 2X7R-5KGZ-Z 016X-52P8-4 041V-48PE-D 0A61-74NU-C 0J71-6UGD-Z 0P4Y-78J7-V 0R1W-2MNL-2 0T41-35J4-A 0V7B-4EJH-A 0W65-4PJB-V 1036-4KHG-N 156Y-70HU-R 1A4K-2BNM-C 1D5N-0YG8-K 1Y2E-0NNG-X 1Z7W-00G0-F 2141-35J4-G 2630-65P4-3 2A5H-1YML-U 2K53-2END-0 2M2G-15LB-E 2N78-6CPV-9 2V3T-67ME-K 2X44-7ELE-J 2X46-12JD-H 2Z73-2NL7-2 302H-6LPP-W 3107-4RG3-6 375F-7UMF-K 3B54-58M8-P 3K33-61K2-0 3Z4R-2LKY-9 483K-28MC-M 4J6V-6XMZ-R 4N5A-1NH8-5 4T7D-4EJH-E 4V52-30PD-F 4V6C-1PJ5-R 521J-18LJ-5 591J-18LJ-G 5A4L-1CJJ-J 5A4T-57JE-R 5C4L-7WMK-2 5F72-2NL7-Y 5G1Y-1WJ2-5 5Y6J-3HL1-E 6031-24GX-X 684V-57JE-Z 693T-6YM2-Y 6D2X-1BLC-4 6F3G-46MR-K 6L0D-02P6-4 6M5J-0JLM-H 6N1K-25G7-8 6U32-24GX-0 6Y79-1HM0-M 6Z1H-1KG1-6 6Z5N-0YG8-R 724E-34MB-4 7U76-1HM0-D 7Y5Z-7FHX-F 4L7J-7MPL-J 0D6E-39G4-P 5B0U-08NR-E 6F2V-1BLC-W 003W-3UN7-M 2V57-6TKK-N 3D5U-2PJT-3 3T5P-0YG8-Y 0N1E-7HJU-5 1J5H-1YML-3 1W6W-21KC-Z 284W-72P0-G 325N-73HZ-G 4U39-3JGK-2 5G2W-1VNP-A 5L5X-32PN-Z 617X-31N5-0 017L-0VKT-B 067L-7MPL-2 116V-21KC-Z 116Z-1ZL8-V 1305-0UH7-M 1K1B-1EGA-D 2P1T-75JN-B 3055-68PX-3 350Y-2DMM-7 3A0H-6AMX-6 3V31-7NPM-F 414L-03P1-E 4R03-1FHE-U 4R1J-18LJ-4 523N-67ME-8 5X3N-67ME-P 623Z-2RHR-9 6633-24GX-D 670F-60JV-1 6L0F-60JV-U 713Z-3UN7-Y 7676-1HM0-9 7C5C-7UMF-D 1135-04K8-Z 094B-40KU-3 2C6N-3KLU-C 2V2W-3VM1-X 545X-32PN-F 706P-0PLF-0 6X1X-3CH6-V 264N-2LKY-K 3C44-7ELE-4 3K0J-6GLT-B 6T49-44NB-Z 2V51-29H4-H 4K30-5BNT-7 5N40-2RHR-N 0Y0L-0FKE-E 3F5L-5AGH-9 6U2G-3YNZ-J 032M-79HL-V 115M-4LGG-4 2L0L-0FKE-A 0D1V-48PE-J 4L7Y-00G0-F 567W-7BG9-5 6E64-7XLV-M 6J3E-0KN0-2 534U-57JE-Y 472X-65P4-L 7T0A-02P6-1 0U28-2WG5-M 0F6V-6XMZ-5 0T1M-18LJ-R 0Y5Z-7FHX-N 1256-68PX-G 1R44-7ELE-2 2W26-0CKW-V 3Z3G-0KN0-0 4H5T-0YG8-P 4R0G-60JV-1 5K0P-17M9-H 6L3G-46MR-4 3F3D-41PZ-F 661M-4HGE-D 5G2A-6DHP-W 6M6N-0DN4-G 121R-22HW-3 7W3Y-0LJ3-B 4W2V-3VM1-7 6R2J-47P7-9 0K6E-39G4-5
Quantity recalled257,183 Units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1386-2024
FDA event ID
94185
Recalling firm
Philips North America, Cambridge, MA
Classification
Class II
Status
Ongoing
Recall initiated
28 Feb 2024
FDA classified
27 Mar 2024
Reported
3 Apr 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

28 Feb 2024Recall initiated by company
27 Mar 2024Classified by FDA+28 days
3 Apr 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Patient Information Center iX (PIC iX), Model Numbers 866389, 866390 (PIC iX Expand), and 866424 (PIC iX Hardware) running Software Version Number 4.X

Where it was distributed

Worldwide - US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.

Which lots are affected?

UDI-DI: (01)00884838104594(10)4.0.1 (01)00884838104594(10)4.0.2 (01)00884838112070(10)4.1.0 (01)00884838112070(10)4.1.1 (01)00884838121782(10)4.2.0 (01)00884838121782(10)4.2.1; Serial Numbers: 7G6G-7PGV-W 5971-2NL7-Z 6D62-19MD-9 463N-67ME-X 5K2A-6DHP-K 5E42-35J4-P 4E6Z-70HU-B 1T12-0MNA-C 2H34-3MMP-U 6P1E-7HJU-W 3R6X-3XNY-6 7F50-53NK-L 3Z3B-1AHM-H 0C5Y-7FHX-V 0X7R-2AGC-Y 0P5D-5VMH-V 6627-6MKH-U 5J7B-1XPT-T 7U4C-44NB-E 2Y7R-5KGZ-W 1P22-0XNW-P 1004-6RJX-E 1U0P-0FKE-C 396T-5ZH1-5 610R-7AHH-2 0B5K-10GT-X 0Z7K-0VKT-0 2A5D-1NH8-Y 2Y5M-5AGH-E 3153-2END-M 4H2B-5YJP-0 5F6T-6XMZ-Y 5Y3Z-2RHR-N 6B6G-3DKJ-4 7Z05-0UH7-V 5H6A-2ZKV-1 1J61-2KH5-M 612A-2WG5-W 737K-7LPC-G 7Z3Y-0LJ3-2 3T0R-7AHH-X 7A4Y-78J7-V 2P5H-0JLM-Y 002A-5MGJ-6 0X63-19MD-G 221G-5NMA-3 4W6U-0PLF-2 7A3M-6VGP-H 7J0C-69LP-H 0V4G-0EG2-B 2K1C-42GL-Y 4M77-1HM0-G 4R5J-5AGH-B 4T62-19MD-7 4W08-0UH7-U 5Y2X-65P4-E 6H5T-3FHA-4 6K59-0HN3-B 2X7R-5KGZ-Z 016X-52P8-4 041V-48PE-D 0A61-74NU-C 0J71-6UGD-Z 0P4Y-78J7-V 0R1W-2MNL-2 0T41-35J4-A 0V7B-4EJH-A 0W65-4PJB-V 1036-4KHG-N 156Y-70HU-R 1A4K-2BNM-C 1D5N-0YG8-K 1Y2E-0NNG-X 1Z7W-00G0-F 2141-35J4-G 2630-65P4-3 2A5H-1YML-U 2K53-2END-0 2M2G-15LB-E 2N78-6CPV-9 2V3T-67ME-K 2X44-7ELE-J 2X46-12JD-H 2Z73-2NL7-2 302H-6LPP-W 3107-4RG3-6 375F-7UMF-K 3B54-58M8-P 3K33-61K2-0 3Z4R-2LKY-9 483K-28MC-M 4J6V-6XMZ-R 4N5A-1NH8-5 4T7D-4EJH-E 4V52-30PD-F 4V6C-1PJ5-R 521J-18LJ-5 591J-18LJ-G 5A4L-1CJJ-J 5A4T-57JE-R 5C4L-7WMK-2 5F72-2NL7-Y 5G1Y-1WJ2-5 5Y6J-3HL1-E 6031-24GX-X 684V-57JE-Z 693T-6YM2-Y 6D2X-1BLC-4 6F3G-46MR-K 6L0D-02P6-4 6M5J-0JLM-H 6N1K-25G7-8 6U32-24GX-0 6Y79-1HM0-M 6Z1H-1KG1-6 6Z5N-0YG8-R 724E-34MB-4 7U76-1HM0-D 7Y5Z-7FHX-F 4L7J-7MPL-J 0D6E-39G4-P 5B0U-08NR-E 6F2V-1BLC-W 003W-3UN7-M 2V57-6TKK-N 3D5U-2PJT-3 3T5P-0YG8-Y 0N1E-7HJU-5 1J5H-1YML-3 1W6W-21KC-Z 284W-72P0-G 325N-73HZ-G 4U39-3JGK-2 5G2W-1VNP-A 5L5X-32PN-Z 617X-31N5-0 017L-0VKT-B 067L-7MPL-2 116V-21KC-Z 116Z-1ZL8-V 1305-0UH7-M 1K1B-1EGA-D 2P1T-75JN-B 3055-68PX-3 350Y-2DMM-7 3A0H-6AMX-6 3V31-7NPM-F 414L-03P1-E 4R03-1FHE-U 4R1J-18LJ-4 523N-67ME-8 5X3N-67ME-P 623Z-2RHR-9 6633-24GX-D 670F-60JV-1 6L0F-60JV-U 713Z-3UN7-Y 7676-1HM0-9 7C5C-7UMF-D 1135-04K8-Z 094B-40KU-3 2C6N-3KLU-C 2V2W-3VM1-X 545X-32PN-F 706P-0PLF-0 6X1X-3CH6-V 264N-2LKY-K 3C44-7ELE-4 3K0J-6GLT-B 6T49-44NB-Z 2V51-29H4-H 4K30-5BNT-7 5N40-2RHR-N 0Y0L-0FKE-E 3F5L-5AGH-9 6U2G-3YNZ-J 032M-79HL-V 115M-4LGG-4 2L0L-0FKE-A 0D1V-48PE-J 4L7Y-00G0-F 567W-7BG9-5 6E64-7XLV-M 6J3E-0KN0-2 534U-57JE-Y 472X-65P4-L 7T0A-02P6-1 0U28-2WG5-M 0F6V-6XMZ-5 0T1M-18LJ-R 0Y5Z-7FHX-N 1256-68PX-G 1R44-7ELE-2 2W26-0CKW-V 3Z3G-0KN0-0 4H5T-0YG8-P 4R0G-60JV-1 5K0P-17M9-H 6L3G-46MR-4 3F3D-41PZ-F 661M-4HGE-D 5G2A-6DHP-W 6M6N-0DN4-G 121R-22HW-3 7W3Y-0LJ3-B 4W2V-3VM1-7 6R2J-47P7-9 0K6E-39G4-5

Where was it sold?

Worldwide - US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 94185

This product is part of a recall event; the main page for the event is CareEvent Event Management System recalled by Philips North America.

More recalls from Philips North America

All 531

Other Medical Software recalls

All 903