CLASS II Device recall · Reported 28 Feb 2018 · FDA Enforcement Report
Pentax Video Duodenoscope recalled by Pentax of America
Pentax of America is recalling Pentax Video Duodenoscope, distributed nationwide in the US. The recall was initiated on 13 Feb 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0643-2018
- FDA event ID
- 79237
- Recalling firm
- Pentax of America, Montvale, NJ
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 13 Feb 2018
- FDA classified
- 20 Feb 2018
- Reported
- 28 Feb 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Processing Deviation
- Device type
- Patient Monitors
Timeline
| 13 Feb 2018 | Recall initiated by company | |
| 20 Feb 2018 | Classified by FDA | +7 days |
| 28 Feb 2018 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The duodenoscopes are being recalled in order to replace the forceps elevator mechanism, the O-rings, and the distal end covering to be consistent with the updated design as well as provide an updated periodic inspection as part of the Operation Manual in order to mitigate the potential risk of infection in flexible endoscopy.
The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.
Product as listed by the FDA
Pentax Video Duodenoscope Model: ED-3490TK (UDI of design being recalled: 04961333232420) These instruments are intended to provide optical visualization of (via a video monitor), and therapeutic access to, the biliary tract via the upper GI tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: esophagus, stomach, duodenum, common bile, hepatic, and cystic ducts.
Where it was distributed
USA (nationwide) Distribution to the states of including Puerto Rico : AL, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, WV and WY, and the District of Columbia.
Nationwide AlabamaArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouri
Questions about this recall
Why was it recalled?
The duodenoscopes are being recalled in order to replace the forceps elevator mechanism, the O-rings, and the distal end covering to be consistent with the updated design as well as provide an updated periodic inspection as part of the Operation Manual in order to mitigate the potential risk of infection in flexible endoscopy.
Which lots are affected?
All Serial Numbers
Where was it sold?
USA (nationwide) Distribution to the states of including Puerto Rico : AL, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, WA, WI, WV and WY, and the District of Columbia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Pentax of America
All 33| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Pentax Medical Video Processor recalled by Pentax of America | Pentax of America | Other | 20 states |
| Device | CLASS II | Pentax Medical Video Colonoscope- Intended to provide recalled | Pentax of America | Other | Nationwide |
| Device | CLASS II | Pentax Medical Video Upper GI Scope-Intended to provide recalled | Pentax of America | Other | Nationwide |
| Device | CLASS II | Pentax Medical Video Processor- Intended to be used recalled | Pentax of America | Other | Nationwide |
| Device | CLASS II | Colonoscope Family # 1-Pentax Video Colonoscopes recalled | Pentax of America | Other | Nationwide |
Other Patient Monitors recalls
All 1,022| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Passport Series Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | T1 Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | cPM Series Patient Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | BeneVision Series Telemetry Monitor recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |
| Device | CLASS III | Panorama and Telemetry Monitoring System recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |