CLASS II Device recall · Reported 17 Dec 2014 · FDA Enforcement Report

Perfecto 2 V Oxygen Concentrator The finished device recalled

Invacare is recalling Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit, distributed nationwide in the US. The recall was initiated on 23 Sep 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesThe defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers)
Quantity recalled5,963

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0577-2015
FDA event ID
69441
Recalling firm
Invacare, Elyria, OH
Classification
Class II
Status
Terminated
Recall initiated
23 Sep 2014
FDA classified
9 Dec 2014
Reported
17 Dec 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

23 Sep 2014Recall initiated by company
9 Dec 2014Classified by FDA+77 days
17 Dec 2014Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.

Where it was distributed

Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.

Nationwide ArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIowaIllinoisIndianaKentuckyLouisianaMassachusettsMichiganMissouriMississippiNorth CarolinaNebraskaNew HampshireNew Jersey

Questions about this recall

Why was it recalled?

It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.

Which lots are affected?

The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers)

Where was it sold?

Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Invacare

All 35

Other CPAP & Ventilators recalls

All 556