CLASS II Device recall · Reported 17 Dec 2014 · FDA Enforcement Report
Perfecto 2 V Oxygen Concentrator The finished device recalled
Invacare is recalling Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit, distributed nationwide in the US. The recall was initiated on 23 Sep 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0577-2015
- FDA event ID
- 69441
- Recalling firm
- Invacare, Elyria, OH
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 23 Sep 2014
- FDA classified
- 9 Dec 2014
- Reported
- 17 Dec 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- CPAP & Ventilators
Timeline
| 23 Sep 2014 | Recall initiated by company | |
| 9 Dec 2014 | Classified by FDA | +77 days |
| 17 Dec 2014 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.
Where it was distributed
Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
Nationwide ArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIowaIllinoisIndianaKentuckyLouisianaMassachusettsMichiganMissouriMississippiNorth CarolinaNebraskaNew HampshireNew Jersey
Questions about this recall
Why was it recalled?
It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.
Which lots are affected?
The defective sieve bed subassembly manufactured from 7/31/2014 to 8/19/2014 impacts the finished device- IRC5PO2V and Service Part Number-1165099. The serial number range is as follows for the finished device: 14GF053772-14HF032124 (Sequential serial numbers)
Where was it sold?
Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Invacare
All 35| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Dolomite Gloss Rollator - intended to support during recalled | Invacare | Foreign Material | Nationwide |
| Device | CLASS III | Platinum 5NXG Oxygen Concentrator recalled by Invacare | Invacare | Other | 18 states |
| Device | CLASS II | TDX SP2 recalled by Invacare | Invacare | Device Malfunction | Nationwide |
| Device | CLASS II | Aviva Storm RX with LiNX Gyro Power Wheelchair recalled by Invacare | Invacare | Other | Nationwide |
| Device | CLASS II | TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module recalled | Invacare | Other | Nationwide |
Other CPAP & Ventilators recalls
All 556| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Multiple-type respiratory pathogen nucleic acid IVD, kit recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Multiple-type respiratory pathogen nucleic acid IVD, kit recalled | BioFire Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide |