CLASS II Device recall · Reported 18 May 2022 · FDA Enforcement Report
Philips Hemodynamic Application recalled by Philips North America
Philips North America is recalling Philips Hemodynamic Application, distributed in Florida, Idaho, New Jersey, Oklahoma, South Carolina, Texas. The recall was initiated on 22 Feb 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1041-2022
- FDA event ID
- 89878
- Recalling firm
- Philips North America, Cambridge, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 22 Feb 2022
- FDA classified
- 6 May 2022
- Reported
- 18 May 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Medical Software
Timeline
| 22 Feb 2022 | Recall initiated by company | |
| 6 May 2022 | Classified by FDA | +73 days |
| 18 May 2022 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
Philips Hemodynamic Application, Model Number 722463, Software Version Number: R1.2.2 and R1.2.3. For visualization and analysis of patient hemodynamic data in the Philips Interventional Hemodynamic System.
Where it was distributed
Devices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.
Questions about this recall
Is Philips Hemodynamic Application recalled?
Yes. Philips North America recalled Philips Hemodynamic Application on 22 Feb 2022. The recall is Class II and its status is Terminated. Recall number Z-1041-2022.
Why was it recalled?
When a Neonatal patient is admitted, the Non-Invasive Blood Pressure is not set to Neonatal mode on the Philips Xperflex Cardio FC2010 patient monitor. If the NIBP is not set to Neonatal, the cuff inflation pressure applied will be higher than 150 mmHg and the deflation pressure will be higher than 5 mmHg. An applied cuff inflation pressure higher than 150mmHg or a deflation pressure higher than 5 could lead to bruising, mmHg nerve damage or bone deformity for a neonate.
Which lots are affected?
UDI: B-00884838084544; System Serial Numbers: 396 411 225 351 51 46 33 189 132 19 51 924 644 645 646 376 384 487 538 886 14 15 643 1194 220 1238 947 949 950 951 89 657 169 168 413 414 624 565 1060 393 512 278 818 362 123 802 395 576 932 291 834 539 605 161 970 686 761 229 230
Where was it sold?
Devices were distributed to the following US states: FL, ID, NJ, OK, SC and TX. Devices were distributed to the following foreign countries: France, Germany, India, Mexico, Netherlands, New Zealand, Norway, Poland, South Africa, and Switzerland.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips North America
All 531Other Medical Software recalls
All 903| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT Product Codes recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT Product Codes recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel Product recalled | Philips North America | Device Malfunction | MA |
| Device | ANNOUNCED | BMC Luna G3 APAP recalled by BMC MedicalBMC | BMC Medical | Device Malfunction | Nationwide |