CLASS II Device recall · Reported 16 Jan 2013 · FDA Enforcement Report
Philips and Laerdal brands of HeartStart HS1 Defibrillator Family recalled
Philips Medical Systems North America Co. Phillips is recalling Philips and Laerdal brands of HeartStart HS1 Defibrillator Family, distributed nationwide in the US. The recall was initiated on 26 Jul 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0643-2013
- FDA event ID
- 63030
- Recalling firm
- Philips Medical Systems North America Co. Phillips, Bothell, WA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 26 Jul 2012
- FDA classified
- 4 Jan 2013
- Reported
- 16 Jan 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Pacemakers & Defibrillators
Timeline
| 26 Jul 2012 | Recall initiated by company | |
| 4 Jan 2013 | Classified by FDA | +162 days |
| 16 Jan 2013 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Philips determined that it is important to clarify information in the Owner's Manual and keep customers informed about the maintenance of their FRx (models 861304 and 861305), OnSite (models M5066A and M5067A), or HeartStart Home (model M5068A) automated external defibrillators (AEDs) shipped between December 2005 and July 2012. Philips had determined there is a need to emphasize the importance o
Product as listed by the FDA
Philips and Laerdal brands of HeartStart HS1 Defibrillator Family Model number M5066A, M5067A, and M5068A Automated External Defibrillator Manufactured by Philips Medical Systems, Seattle, WA 98121 USA. The HS1 Defibrillator is intended to treat ventricular fibrillation, the most common cause of sudden cardiac arrest. Using voice prompts, light emitting diodes (LEDs) and buttons, the user is guided through the response. The HS1 uses a SMART biphasic, impedance compensating exponential waveform to deliver a nominal 150 J to adults and nominal 50 J to infants/children.
Where it was distributed
Worldwide Distribution - USA (nationwide) and the countries of Japan and Malaysia.
Questions about this recall
Why was it recalled?
Philips determined that it is important to clarify information in the Owner's Manual and keep customers informed about the maintenance of their FRx (models 861304 and 861305), OnSite (models M5066A and M5067A), or HeartStart Home (model M5068A) automated external defibrillators (AEDs) shipped between December 2005 and July 2012. Philips had determined there is a need to emphasize the importance o
Which lots are affected?
HS1 product family all start with A in the serial number and break down into 3 catalog numbers - M5066A, M5067A and M5068A: 1) M5066A (referred to as Onsite or HS1), Philips brand, 510 (k) = K020715, sold nationally and internationally, serial numbers range from A021-00038 through A12E-03490, with 306,108 total devices. 2) M5067A, Philips brand 510(k) = K040904, sold nationally and internationally, serial numbers range from A05C-00727 through A12D-01406, with 978 total devices. 3) M5068A, Philips brand 510(k) = K020715, sold nationally and internationally, serial numbers range from A021J-00207 through A12E-01900, with 22,760 total devices 4) M5066A, Laerdal brand, 510 (k) = K020715, sold internationally, serial numbers range from A021-00036 through A12D-02148, with 60,433 total devices. 5) M5067A, Laerdal brand, sold as Home device- cleared for OTC on initial 510(k) = K040904, serial numbers range from A021-00133 through A11B-00970, with 8659 total devices. 6) M5068A, Laerdal brand, 510(k) = K020715, sold internationally, serial numbers range from A02K-00172 through A11H-00868, with 5217 total devices. Please note the serial number range represents first and last serial number for each model. Serial numbers are not continguous. Laerdal is a distribution partner in EU and Canada, sold internationally
Where was it sold?
Worldwide Distribution - USA (nationwide) and the countries of Japan and Malaysia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 63030This product is part of a recall event; the main page for the event is Philips HeartStart FRx 861304 Automated External Defibrillator recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Philips HeartStart FRx 861304 Automated External Defibrillator recalled | Philips Medical Systems North America Co. Phillips | Other | Nationwide |
More recalls from Philips Medical Systems North America Co. Phillips
All 2| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips HeartStart FRx 861304 Automated External Defibrillator recalled | Philips Medical Systems North America Co. Phillips | Other | Nationwide |
Other Pacemakers & Defibrillators recalls
All 494| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DermaV Laser System recalled by Lutronic | Lutronic | Other | Nationwide |
| Device | CLASS II | Convenience kits containing select SKUs of Polycarbonate recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits: Loop Pack DYNJ61460C Open Heart recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Medline convenience kits: Acl Pack DYNJ64161C Arthroscopy recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Medline convenience kits: Bib AV Fistula Pack DYNJ58304B recalled | Medline Industries | Device Malfunction | Nationwide |