CLASS II ONGOING Device recall · Reported 7 Oct 2026 · FDA Enforcement Report

Philips Medical Systems Technologies Ltd recalled

Philips North America is recalling Philips Medical Systems Technologies Ltd. The recall was initiated on 18 Aug 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: (01)00884838101111
Quantity recalled62

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3155-2026
FDA event ID
99865
Recalling firm
Philips North America, Cambridge, MA
Classification
Class II
Status
Ongoing
Recall initiated
18 Aug 2026
FDA classified
29 Sep 2026
Reported
7 Oct 2026
Type
FDA Mandated
Reason category
Device Malfunction
Device type
Medical Software

Timeline

18 Aug 2026Recall initiated by company
29 Sep 2026Classified by FDA+42 days
7 Oct 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Philips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333

Where it was distributed

U.S. and U.S. Territories.

Questions about this recall

Why was it recalled?

The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.

Which lots are affected?

UDI: (01)00884838101111

Where was it sold?

U.S. and U.S. Territories.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 13 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Philips North America

All 558

Other Medical Software recalls