CLASS II ONGOING Device recall · Reported 22 Apr 2026 · FDA Enforcement Report

Philips Spectral CT on Rails recalled by Philips Medical Systems

Philips Medical Systems is recalling Philips Spectral CT on Rails, distributed nationwide in the US. The recall was initiated on 7 Mar 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.
Quantity recalled3 systems

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1825-2026
FDA event ID
98588
Recalling firm
Philips Medical Systems, Cambridge, MA
Classification
Class II
Status
Ongoing
Recall initiated
7 Mar 2026
FDA classified
16 Apr 2026
Reported
22 Apr 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Mar 2026Recall initiated by company
16 Apr 2026Classified by FDA+40 days
22 Apr 2026Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Philips Spectral CT on Rails. Model Number: 728334.

Where it was distributed

Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.

Nationwide Minnesota

Questions about this recall

Why was it recalled?

Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing Go. 2. Potential where the message of Previous Surview Exists Select Previous Surview? <Yes> or <No> should display but does not appear. 3. After performing Surview and planning the Brain Helical acquisition by setting Brain Area DoseRight Index to increase the dose, there is the potential that the Define Head Area option in the context menu is grayed out.

Which lots are affected?

Model Number: 728334. UDI: (01)00884838103627(21)1005, (01)00884838103627(21)1006, (01)00884838103627(21)1007. Software Version Number: 5.1.X. Serial Numbers: 1005, 1006, 1007.

Where was it sold?

Worldwide - US Nationwide distribution in the states of Minnesota and the countries of France and Netherlands.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Philips Medical Systems

All 82

Other Other Medical Devices recalls

All 13,908