CLASS II Device recall · Reported 24 Mar 2021 · FDA Enforcement Report

Philips CT systems are advanced continuous-rotation recalled

Philips Healthcare (Suzhou) is recalling Philips CT systems are advanced continuous-rotation computed tomography systems suitable, distributed nationwide in the US. The recall was initiated on 24 Feb 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUnique Device Identifier (01)00884838059849(21)304034 (01)00884838059849(21)325078 (01)00884838085015(21)500218 (01)00884838085015(21)500219 (01)00884838085015(21)500221 (01)00884838085015(21)500222 (01)00884838085015(21)500223 (01)00884838085015(21)500224 (01)00884838085015(21)500225 (01)00884838085015(21)500226 (01)00884838085015(21)500227 (01)00884838085015(21)500229 (01)00884838085015(21)500231 (01)00884838085015(21)500233 (01)00884838085015(21)500234 (01)00884838085015(21)500237 (01)00884838085015(21)500238 (01)00884838085015(21)500239 (01)00884838085015(21)500241 (01)00884838085015(21)500242 (01)00884838085015(21)500245 (01)00884838085015(21)500247 (01)00884838085015(21)500248 (01)00884838085015(21)500249 (01)00884838085015(21)500251 (01)00884838085015(21)500252 (01)00884838085015(21)500254 (01)00884838085015(21)500255 (01)00884838085015(21)500256 (01)00884838085015(21)500257 (01)00884838085015(21)500258 (01)00884838085015(21)500261 (01)00884838085015(21)500264 (01)00884838085015(21)500274 (01)00884838085015(21)500283 (01)00884838085015(21)500296 (01)00884838085015(21)530038 (01)00884838085015(21)530067 (01)00884838085015(21)530071 (01)00884838085015(21)530098 (01)00884838085015(21)530099 (01)00884838085015(21)530101 (01)00884838085015(21)530103 (01)00884838085015(21)530104 (01)00884838085015(21)530106 (01)00884838085015(21)530107 (01)00884838085015(21)530108 (01)00884838085015(21)530109 (01)00884838085015(21)530111 (01)00884838085015(21)530112 (01)00884838085015(21)530113 (01)00884838085015(21)530114 (01)00884838085015(21)530115 (01)00884838085015(21)530116 (01)00884838085015(21)530117 (01)00884838085015(21)530118 (01)00884838085015(21)530119 (01)00884838085015(21)530120 (01)00884838085015(21)530121 (01)00884838085015(21)530124 (01)00884838085015(21)530126 (01)00884838085015(21)530127 (01)00884838085015(21)530128 (01)00884838085015(21)530129 (01)00884838085015(21)530130 (01)00884838085015(21)530132 (01)00884838085015(21)530133 (01)00884838085015(21)530134 (01)00884838085015(21)530135 (01)00884838085015(21)530136 (01)00884838085015(21)530137 (01)00884838085015(21)530138 (01)00884838085015(21)530141 (01)00884838085015(21)530148 (01)00884838085015(21)530150 (01)00884838085015(21)530152 (01)00884838085015(21)530156 (01)00884838085015(21)530157 (01)00884838085015(21)530159 (01)00884838085015(21)530160 (01)00884838085015(21)530163 (01)00884838085015(21)530165 (01)00884838085015(21)530166 (01)00884838085015(21)530167 (01)00884838085015(21)530169 (01)00884838085015(21)530170 (01)00884838085015(21)530171 (01)00884838085015(21)530172 (01)00884838085015(21)530173 (01)00884838085015(21)530174 (01)00884838085015(21)530175 (01)00884838085015(21)530176 (01)00884838085015(21)530177 (01)00884838085015(21)530178 (01)00884838085015(21)530180 (01)00884838085015(21)530181 (01)00884838085015(21)530182 (01)00884838085015(21)530183 (01)00884838085015(21)530186 (01)00884838085015(21)530187 (01)00884838085015(21)530188 (01)00884838085015(21)530189 (01)00884838085015(21)530190 (01)00884838085015(21)530191 (01)00884838085015(21)530192 (01)00884838085015(21)530194 (01)00884838085015(21)530197 (01)00884838085015(21)530198 (01)00884838085015(21)500230
Quantity recalled109 systems

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1273-2021
FDA event ID
87473
Recalling firm
Philips Healthcare (Suzhou), Suzhou, N/A
Classification
Class II
Status
Terminated
Recall initiated
24 Feb 2021
FDA classified
17 Mar 2021
Reported
24 Mar 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

24 Feb 2021Recall initiated by company
17 Mar 2021Classified by FDA+21 days
24 Mar 2021Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Philips CT systems are advanced continuous-rotation computed tomography systems suitable for a wide range of computed tomographic (CT) applications - Product Usage: intended for use in the head and whole body.

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of KY, FL, St. Croix, and the countries of Australia, Bolivia, China, Germany, Hungary, Italy, Japan, Republic of Korea, Latvia, Palestine, Puerto Rico, Romania, Russia, Saudi Arabia, Senegal, Serbia, Spain, Sweden, Thailand, Uzbekistan.

Nationwide FloridaKentuckyPuerto Rico

Questions about this recall

Why was it recalled?

Philips has internally detected an issue with the actuators used in the Incisive CT couch, which could result in the couch failure to self-lock, and unexpectedly moving slowly downward to its limit position.

Which lots are affected?

Unique Device Identifier (01)00884838059849(21)304034 (01)00884838059849(21)325078 (01)00884838085015(21)500218 (01)00884838085015(21)500219 (01)00884838085015(21)500221 (01)00884838085015(21)500222 (01)00884838085015(21)500223 (01)00884838085015(21)500224 (01)00884838085015(21)500225 (01)00884838085015(21)500226 (01)00884838085015(21)500227 (01)00884838085015(21)500229 (01)00884838085015(21)500231 (01)00884838085015(21)500233 (01)00884838085015(21)500234 (01)00884838085015(21)500237 (01)00884838085015(21)500238 (01)00884838085015(21)500239 (01)00884838085015(21)500241 (01)00884838085015(21)500242 (01)00884838085015(21)500245 (01)00884838085015(21)500247 (01)00884838085015(21)500248 (01)00884838085015(21)500249 (01)00884838085015(21)500251 (01)00884838085015(21)500252 (01)00884838085015(21)500254 (01)00884838085015(21)500255 (01)00884838085015(21)500256 (01)00884838085015(21)500257 (01)00884838085015(21)500258 (01)00884838085015(21)500261 (01)00884838085015(21)500264 (01)00884838085015(21)500274 (01)00884838085015(21)500283 (01)00884838085015(21)500296 (01)00884838085015(21)530038 (01)00884838085015(21)530067 (01)00884838085015(21)530071 (01)00884838085015(21)530098 (01)00884838085015(21)530099 (01)00884838085015(21)530101 (01)00884838085015(21)530103 (01)00884838085015(21)530104 (01)00884838085015(21)530106 (01)00884838085015(21)530107 (01)00884838085015(21)530108 (01)00884838085015(21)530109 (01)00884838085015(21)530111 (01)00884838085015(21)530112 (01)00884838085015(21)530113 (01)00884838085015(21)530114 (01)00884838085015(21)530115 (01)00884838085015(21)530116 (01)00884838085015(21)530117 (01)00884838085015(21)530118 (01)00884838085015(21)530119 (01)00884838085015(21)530120 (01)00884838085015(21)530121 (01)00884838085015(21)530124 (01)00884838085015(21)530126 (01)00884838085015(21)530127 (01)00884838085015(21)530128 (01)00884838085015(21)530129 (01)00884838085015(21)530130 (01)00884838085015(21)530132 (01)00884838085015(21)530133 (01)00884838085015(21)530134 (01)00884838085015(21)530135 (01)00884838085015(21)530136 (01)00884838085015(21)530137 (01)00884838085015(21)530138 (01)00884838085015(21)530141 (01)00884838085015(21)530148 (01)00884838085015(21)530150 (01)00884838085015(21)530152 (01)00884838085015(21)530156 (01)00884838085015(21)530157 (01)00884838085015(21)530159 (01)00884838085015(21)530160 (01)00884838085015(21)530163 (01)00884838085015(21)530165 (01)00884838085015(21)530166 (01)00884838085015(21)530167 (01)00884838085015(21)530169 (01)00884838085015(21)530170 (01)00884838085015(21)530171 (01)00884838085015(21)530172 (01)00884838085015(21)530173 (01)00884838085015(21)530174 (01)00884838085015(21)530175 (01)00884838085015(21)530176 (01)00884838085015(21)530177 (01)00884838085015(21)530178 (01)00884838085015(21)530180 (01)00884838085015(21)530181 (01)00884838085015(21)530182 (01)00884838085015(21)530183 (01)00884838085015(21)530186 (01)00884838085015(21)530187 (01)00884838085015(21)530188 (01)00884838085015(21)530189 (01)00884838085015(21)530190 (01)00884838085015(21)530191 (01)00884838085015(21)530192 (01)00884838085015(21)530194 (01)00884838085015(21)530197 (01)00884838085015(21)530198 (01)00884838085015(21)500230

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of KY, FL, St. Croix, and the countries of Australia, Bolivia, China, Germany, Hungary, Italy, Japan, Republic of Korea, Latvia, Palestine, Puerto Rico, Romania, Russia, Saudi Arabia, Senegal, Serbia, Spain, Sweden, Thailand, Uzbekistan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Philips Healthcare (Suzhou)

All 4

Other Other Medical Devices recalls

All 13,857