CLASS II Device recall · Reported 24 Dec 2014 · FDA Enforcement Report
Pkg Insert, Atraumatic Grasper Forceps, Double Action recalled
Stryker Endoscopy is recalling Pkg Insert, Atraumatic Grasper Forceps, Double Action, distributed nationwide in the US. The recall was initiated on 17 Nov 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0710-2015
- FDA event ID
- 69838
- Recalling firm
- Stryker Endoscopy, San Jose, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 17 Nov 2014
- FDA classified
- 17 Dec 2014
- Reported
- 24 Dec 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Heart & Vascular Devices
Timeline
| 17 Nov 2014 | Recall initiated by company | |
| 17 Dec 2014 | Classified by FDA | +30 days |
| 24 Dec 2014 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
PKG, 5MM INSERT, ATRAUMATIC GRASPER FORCEPS, DOUBLE ACTION 45CM, P/N 0250080585 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Where it was distributed
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Questions about this recall
Is Pkg Insert, Atraumatic Grasper Forceps, Double Action recalled?
Yes. Stryker Endoscopy recalled Pkg Insert, Atraumatic Grasper Forceps, Double Action on 17 Nov 2014. The recall is Class II and its status is Terminated. Recall number Z-0710-2015.
Why was it recalled?
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Which lots are affected?
343383 941246 941725 942278 942852 943499 944696 945720 1041625 1042488 1042730 1042738 1042827 1140149 1141793 1142618 1143650 1144500 1145003 1145229 1146369 1146656 1242931 1243555 1244994 1245515 1245818 1340539 1341533 1341554 1342765 1343189 1343300 1343410 1343707 1345828 1346105 1346547 1346654 1440433 1442515 1443224 1443605
Where was it sold?
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Stryker Endoscopy
All 238| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | PneumoSure XL High Flow Insufflator recalled by Stryker Endoscopy | Stryker Endoscopy | Other | Nationwide |
| Device | CLASS II | Serfas 90 degree Energy Probe recalled by Stryker Endoscopy | Stryker Endoscopy | Device Malfunction | Nationwide |
| Device | CLASS II | Pkg, Pennington Forceps recalled by Stryker Endoscopy | Stryker Endoscopy | Sterility | Nationwide |
| Device | CLASS II | Pkg Peek Handle recalled by Stryker Endoscopy | Stryker Endoscopy | Sterility | Nationwide |
| Device | CLASS II | Pkg Insert, Johann Forceps recalled by Stryker Endoscopy | Stryker Endoscopy | Sterility | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |