CLASS II Device recall · Reported 4 May 2016 · FDA Enforcement Report
PneumoSure XL High Flow Insufflator recalled by Stryker Endoscopy
Stryker Endoscopy is recalling PneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator, distributed nationwide in the US. The recall was initiated on 28 Mar 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1566-2016
- FDA event ID
- 73674
- Recalling firm
- Stryker Endoscopy, San Jose, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 28 Mar 2016
- FDA classified
- 25 Apr 2016
- Reported
- 4 May 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Surgical Instruments
Timeline
| 28 Mar 2016 | Recall initiated by company | |
| 25 Apr 2016 | Classified by FDA | +28 days |
| 4 May 2016 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
When operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would appear on the display, immediately suspending insufflation.
Product as listed by the FDA
PneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator; High pressure units (HPU) (40L and 45L). Model Numbers: 620-040-610, 620-040-611, 620-040-613, 620-040-614, 620-040-600, 620-040-601, 620-040-602, 620-040-604, 105-210-684 The Pneumo Sure XL High Flow Insufflator is a device that serves to create a cavity by insufflating CO2 during diagnostics and/or therapeutic laparoscopy. The insufflator consists of a console, that is intended to be connected to a CO2 gas source, and has a tubeset inserted to transport CO2 gas from the insufflator to the connecting trocar.
Where it was distributed
Worldwide distribution. US nationwide including Puerto Rico, China, Hong Kong, India, Japan, Korea, Romania, Vietnam, Singapore, Taiwan, Netherlands, Peru, Australia, Canada, Mexico, Argentina, Chile, Poland, South Africa, and Guatemala.
Questions about this recall
Why was it recalled?
When operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would appear on the display, immediately suspending insufflation.
Which lots are affected?
HPUs affected lot numbers 0005051337-000505052795; PneumoSure insufflators affected serial numbers 1510CE0342-1601CE0581
Where was it sold?
Worldwide distribution. US nationwide including Puerto Rico, China, Hong Kong, India, Japan, Korea, Romania, Vietnam, Singapore, Taiwan, Netherlands, Peru, Australia, Canada, Mexico, Argentina, Chile, Poland, South Africa, and Guatemala.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Stryker Endoscopy
All 238| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Serfas 90 degree Energy Probe recalled by Stryker Endoscopy | Stryker Endoscopy | Device Malfunction | Nationwide |
| Device | CLASS II | Pkg, Pennington Forceps recalled by Stryker Endoscopy | Stryker Endoscopy | Sterility | Nationwide |
| Device | CLASS II | Pkg Peek Handle recalled by Stryker Endoscopy | Stryker Endoscopy | Sterility | Nationwide |
| Device | CLASS II | Pkg Insert, Johann Forceps recalled by Stryker Endoscopy | Stryker Endoscopy | Sterility | Nationwide |
| Device | CLASS II | Pkg, Rotating Handle, Ratchet, Insulated Shaft recalled | Stryker Endoscopy | Sterility | Nationwide |
Other Surgical Instruments recalls
All 3,816| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | REF:2B5LT Ethicon Endo-Surgery recalled by Stryker Sustainability | Stryker Sustainability Solutions | Device Malfunction | Nationwide |
| Device | CLASS II | REF:B11LT Ethicon Endo-Surgery, Inc. Xcel Bladeless Trocar recalled | Stryker Sustainability Solutions | Device Malfunction | Nationwide |
| Device | CLASS II | REF:B12LT Ethicon Endo-Surgery, Inc. Xcel Bladeless Trocar recalled | Stryker Sustainability Solutions | Device Malfunction | Nationwide |
| Device | CLASS II | Hemospray Endoscopic Hemostat recalled by Wilson-Cook Medical | Wilson-Cook Medical | Device Malfunction | Nationwide |
| Device | CLASS II | REF:B5LT Ethicon Endo-Surgery recalled by Stryker Sustainability | Stryker Sustainability Solutions | Device Malfunction | Nationwide |