CLASS II Device recall · Reported 8 Jan 2020 · FDA Enforcement Report
The Plex Elite 9000, a Swept-Source Oct [Ss-oct] recalled
Carl Zeiss Meditec is recalling The Plex Elite 9000, a Swept-Source Oct [Ss-oct], distributed nationwide in the US. The recall was initiated on 16 Oct 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0770-2020
- FDA event ID
- 83229
- Recalling firm
- Carl Zeiss Meditec, Dublin, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 16 Oct 2018
- FDA classified
- 2 Jan 2020
- Reported
- 8 Jan 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Heart & Vascular Devices
Timeline
| 16 Oct 2018 | Recall initiated by company | |
| 2 Jan 2020 | Classified by FDA | +443 days |
| 8 Jan 2020 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
The PLEX Elite 9000, a Swept-Source OCT [SS-OCT] - Product Usage: The PLEX Elite 9000 Swept-Source OCT [SS-OCT] is a non- contact, high resolution, wide field of view tomographic and biomicroscopic imaging device intended for in-vivo viewing, axial cross-sectional and three-dimensional imaging of posterior ocular structures. The device is indicated for visualizing posterior ocular structures including, but not limited to, retina, retinal nerve fiber layer, ganglion cell plus inner plexiform layer, macula, optic nerve head, vitreous and choroid. The PLEX Elite SS-OCT angiography is indicated as an aid in the visualization of vascular structures of the retina and choroid.
Where it was distributed
Worldwide distribution - US Nationwide distribution in the states of CA, TX, NY, MA, OH, PA, IA, , KY, FL, MI, IL, MD and Germany, Japan, Switzerland, France Italy, Australia, Austria, Korea, China, India, Canada, Singapore, Norway, Taiwan, Israel, UAE.
Nationwide CaliforniaFloridaIowaIllinoisKentuckyMassachusettsMarylandMichiganNew YorkOhioPennsylvaniaTexas
Questions about this recall
Why was it recalled?
Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.
Which lots are affected?
OBO /HJO Serial Numbers: PE9000-001, PE9000-002, PE9000-003, PE9000-003,PE9000-004, PE9000-005, PE9000-006, PE9000-007, PE9000-008, PE9000-009, PE9000-0010, PE9000-0011, PE9000-0012, PE9000-0013, PE9000-0014, PE9000-0015, PE9000-0016, PE9000-0020, PE9000-0021, PE9000-0022, PE9000-0023, PE9000-0024, PE9000-0025, PE9000-0026, PE9000-0027, PE9000-0029, PE9000-0030, PE9000-0031, PE9000-0032, PE9000-0033, PE9000-0034, PE9000-0035, PE9000-0036, PE9000-0038, PE9000-0039, PE9000-0040, PE9000-0041, PE9000-0042, PE9000-0043, PE9000-0044, PE9000-0045, PE9000-0046, PE9000-0047, PE9000-0048, PE9000-0049, PE9000-0050, PE9000-0051, PE9000-0052, PE9000-0053, PE9000-0054, PE9000-0055, PE9000-0056, PE9000-0057, PE9000-0058, PE9000-0059, PE9000-0060, PE9000-0061, PE9000-0062, PE9000-0063, PE9000-0064, PE9000-0065, PE9000-0066, PE9000-0067, PE9000-0068, PE9000-0069, PE9000-0070, PE9000-0071, PE9000-0072, PE9000-0073, PE9000-0074, PE9000-0075, PE9000-0076, PE9000-0077, PE9000-0078, PE9000-0079, PE9000-0080, PE9000-0081, PE9000-0082, PE9000-0083, PE9000-0084, PE9000-0085, PE9000-0086, PE9000-0087, PE9000-0088, PE9000-0089, PE9000-0090, PE9000-0091, PE9000-0092, PE9000-0093, PE9000-0094, PE9000-0095, PE9000-0096, PE9000-0097, PE9000-0098, PE9000-0099, PE9000-0100, PE9000-0101, PE9000-0102, PE9000-0103, PE9000-0104, PE9000-0105, PE9000-0106, PE9000-0107, PE9000-0108, PE9000-0109, PE9000-0110, PE9000-0111, PE9000-0112, PE9000-0113, PE9000-0114, PE9000-0115, PE9000-0116, PE9000-0117, PE9000-0118, PE9000-0119, PE9000-0120, PE9000-0121, PE9000-0122, PE9000-0123, PE9000-0124, PE9000-0125, PE9000-0126, PE9000-0127, PE9000-0128, PE9000-0129, PE9000-0130, PE9000-0131, PE9000-0132, PE9000-0133, PE9000-0134, PE9000-0135, PE9000-0137, PE9000-0138, PE9000-0139, PE9000-0140, PE9000-0141, PE9000-0142, PE9000-0143, PE9000-0144, PE9000-0145, PE9000-0146, PE9000-0147, PE9000-0148, PE9000-0149, PE9000-0150, PE9000-0151, PE9000-0152, PE9000-0153, PE9000-0154, PE9000-0155, PE9000-0156, PE9000-0158
Where was it sold?
Worldwide distribution - US Nationwide distribution in the states of CA, TX, NY, MA, OH, PA, IA, , KY, FL, MI, IL, MD and Germany, Japan, Switzerland, France Italy, Australia, Austria, Korea, China, India, Canada, Singapore, Norway, Taiwan, Israel, UAE.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Carl Zeiss Meditec
All 21| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Zeiss Intrabeam Spherical Applicator recalled by Carl Zeiss Meditec AG | Carl Zeiss Meditec | Other | Nationwide |
| Device | CLASS II | Zeiss miLOOP Lens Fragmentation Device recalled by Carl Zeiss Meditec AG | Carl Zeiss Meditec | Other | 32 states |
| Device | CLASS II | Cirrus Hd-oct recalled by Carl Zeiss Meditec | Carl Zeiss Meditec | Other | Nationwide |
| Device | CLASS II | Chassis Label - "Cirrus Hd-oct Rx-Only" Product Label recalled | Carl Zeiss Meditec | Device Malfunction | 52 states |
| Device | CLASS II | IOLMaster recalled by Carl Zeiss Meditec AG | Carl Zeiss Meditec | Device Malfunction | 39 states |
Other Heart & Vascular Devices recalls
All 6,270| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Impella Connect Optical AIC, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Impella Optical Controller, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Automated Impella Controller, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS II | Direxion Torqueable Microcatheter, Ngmc/Fathom/027/Straight/1RO/130 recalled | Boston Scientific Cork | Device Malfunction | Nationwide |
| Device | CLASS II | Direxion Torqueable Microcatheter, Ngmc/Single/021/Straight/1RO/130 recalled | Boston Scientific Cork | Device Malfunction | Nationwide |