CLASS II Device recall · Reported 11 Jun 2014 · FDA Enforcement Report
Plum XL, Micro Macro Plum XL3 with data port recalled by Hospira
Hospira is recalling Plum XL, Micro Macro Plum XL3 with data port, and Micro Macro Plum XL with data port, distributed nationwide in the US. The recall was initiated on 24 Jan 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1703-2014
- FDA event ID
- 67453
- Recalling firm
- Hospira, Lake Forest, IL
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 24 Jan 2014
- FDA classified
- 3 Jun 2014
- Reported
- 11 Jun 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Infusion Pumps
Timeline
| 24 Jan 2014 | Recall initiated by company | |
| 3 Jun 2014 | Classified by FDA | +130 days |
| 11 Jun 2014 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
There is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door cannot appropriately lock the cassette in the right position. Depending on the conditions of a broken door roller pin, a number of events can occur. The door cannot be closed or a cassette/door alarm or a false occlusion
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Plum XL, Micro Macro Plum XL3 with data port, and Micro Macro Plum XL with data port infusion pumps, indicated for intravenous, arterial, short-term epidural, and parenteral administration of general I.V. fluids, medications, nutritional fluids, and blood/blood products to patients in hospital and home care environments.
Where it was distributed
Worldwide Distribution - USA nationwide including District of Columbia, Puerto Rico, and US Virgin Islands; Australia, Argentina, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Israel, Italy, Jordan, Kuwait, Luxembourg, Mexico, Oman, Peru, Poland, Portugal, Qatar, Russia, San Marino, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Yemen.
Questions about this recall
Is Plum XL, Micro Macro Plum XL3 with data port, and Micro Macro Plum XL with data port recalled?
Yes. Hospira recalled Plum XL, Micro Macro Plum XL3 with data port, and Micro Macro Plum XL with data port on 24 Jan 2014. The recall is Class II and its status is Terminated. Recall number Z-1703-2014.
Why was it recalled?
There is the potential for the door roller assembly on the Plum LifeCare 5000 Series and Plum XL families of infusers to break. In the event of a broken door roller pin, the door cannot appropriately lock the cassette in the right position. Depending on the conditions of a broken door roller pin, a number of events can occur. The door cannot be closed or a cassette/door alarm or a false occlusion
Which lots are affected?
List Number: 11555; Serial Numbers: All. List Number: 11855; Serial Numbers: All. List Number: 11859; Serial Numbers: All.
Where was it sold?
Worldwide Distribution - USA nationwide including District of Columbia, Puerto Rico, and US Virgin Islands; Australia, Argentina, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Israel, Italy, Jordan, Kuwait, Luxembourg, Mexico, Oman, Peru, Poland, Portugal, Qatar, Russia, San Marino, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Yemen.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Hospira
All 225| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Drug | CLASS I | Infant 25% Dextrose Injection recalled by HospiraDextrose | Hospira | Particulate Matter | Nationwide |
| Device | CLASS II | Minibore Extension Set with Spin Lock Collar, Non-DEHP, Latex-Free recalled | Hospira | Other | Nationwide |
| Drug | CLASS III | 5% Lidocaine HCL and 5% Dextrose Injection, USP Single-dose ampule recalled | Hospira | Potency & Assay | PR |
| Drug | CLASS II | Fentanyl Citrate Injection, USP, 100 mcg Fentanyl recalled by Hospira | Hospira | Sterility | Nationwide |
| Device | CLASS II | LifeCare PCA 3, PCA recalled by Hospira | Hospira | Device Malfunction | Nationwide |
Other Infusion Pumps recalls
All 546| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Baxter Novum IQ Syringe Pump, Loaner recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS I | Baxter Duo-Vent Solution Set and Male Luer Lock Adapter, Product recalled | Baxter Healthcare | Other | Nationwide |
| Device | CLASS II | Plum Solo Precision IV Pump, v1.0.3 recalled by ICU Medical | ICU Medical | Other | Nationwide |
| Device | CLASS II | Intera Oncology Intera 3000 Hepatic Artery Infusion Pump recalled | Boston Scientific | Packaging Defects | Nationwide |