CLASS II Device recall · Reported 26 Aug 2020 · FDA Enforcement Report

Positron Emission Tomography and Computed Tomography System recalled

Shanghai United Imaging Healthcare is recalling Positron Emission Tomography and Computed Tomography System, distributed nationwide in the US. The recall was initiated on 26 Mar 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 06971576832026, Software Version: R001.3.0.0.750505, Serial Numbers: 200017, 200023, 200024
Quantity recalled3

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2819-2020
FDA event ID
85453
Recalling firm
Shanghai United Imaging Healthcare, Shanghai, N/A
Classification
Class II
Status
Terminated
Recall initiated
26 Mar 2020
FDA classified
18 Aug 2020
Reported
26 Aug 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

26 Mar 2020Recall initiated by company
18 Aug 2020Classified by FDA+145 days
26 Aug 2020Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Positron Emission Tomography and Computed Tomography System, Model: uMI 550, Material Number: 88000057 - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Where it was distributed

US Nationwide distribution including in the state of TX.

Nationwide Texas

Questions about this recall

Why was it recalled?

In affected software version, of Positron emission tomography and computed tomography system, the CT Hounsfield unit curve may not display in real time when using the bolus tracking protocol with the tracker scan function. If this occurs, the subsequent clinical scanning protocols will need to be started manually, which may effect image enhancement, and may result in patient rescan.

Which lots are affected?

UDI: 06971576832026, Software Version: R001.3.0.0.750505, Serial Numbers: 200017, 200023, 200024

Where was it sold?

US Nationwide distribution including in the state of TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Shanghai United Imaging Healthcare

All 33

Other Other Medical Devices recalls

All 13,884