CLASS II ONGOING Device recall · Reported 19 Oct 2022 · FDA Enforcement Report

Positron Emission Tomography and Computed Tomography System recalled

Shanghai United Imaging Healthcare is recalling Positron Emission Tomography and Computed Tomography System, distributed in California, Michigan, Texas, Wisconsin. The recall was initiated on 30 May 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI/Serial Numbers: 06971576832026/230002, 230003, 230004, 230005, 230006, 230007, 230008, 230009, 230010, 230011, 230012, 230013, 230014, 230015, 230016, 230017, 230019, 230020, 230021, 230022, 230023, 230024, 230025, 230026, 230027, 230028, 230029, 230030, 230031, 230032, 230033, 232001
Quantity recalled32

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0087-2023
FDA event ID
90888
Recalling firm
Shanghai United Imaging Healthcare, Shanghai, N/A
Classification
Class II
Status
Ongoing
Recall initiated
30 May 2022
FDA classified
13 Oct 2022
Reported
19 Oct 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

30 May 2022Recall initiated by company
13 Oct 2022Classified by FDA+136 days
19 Oct 2022Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Where it was distributed

Distribution to US states of TX, WI, CA, and MI

CaliforniaMichiganTexasWisconsin

Questions about this recall

Why was it recalled?

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

Which lots are affected?

UDI-DI/Serial Numbers: 06971576832026/230002, 230003, 230004, 230005, 230006, 230007, 230008, 230009, 230010, 230011, 230012, 230013, 230014, 230015, 230016, 230017, 230019, 230020, 230021, 230022, 230023, 230024, 230025, 230026, 230027, 230028, 230029, 230030, 230031, 230032, 230033, 232001

Where was it sold?

Distribution to US states of TX, WI, CA, and MI

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Shanghai United Imaging Healthcare

All 33

Other Other Medical Devices recalls

All 13,884