CLASS II ONGOING Device recall · Reported 3 Sep 2025 · FDA Enforcement Report

PrisMax V3 control Unit-US, Hemodialysis Delivery System recalled

Vantive US Healthcare is recalling PrisMax V3 control Unit-US, Hemodialysis Delivery System. The recall was initiated on 28 Jul 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI 07332414126018, All serial numbers
Quantity recalled2449 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2438-2025
FDA event ID
97233
Recalling firm
Vantive US Healthcare, Deerfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
28 Jul 2025
FDA classified
26 Aug 2025
Reported
3 Sep 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

28 Jul 2025Recall initiated by company
26 Aug 2025Classified by FDA+29 days
3 Sep 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD Self-Test Failure

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

PrisMax V3 control Unit-US, Hemodialysis Delivery System, Product Code 955724

Where it was distributed

Worldwide

Questions about this recall

Why was it recalled?

Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD Self-Test Failure

Which lots are affected?

UDI 07332414126018, All serial numbers

Where was it sold?

Worldwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 97233

This product is part of a recall event; the main page for the event is PrisMax V2-US, Hemodialysis Delivery System recalled.

More recalls from Vantive US Healthcare

All 12

Other Other Medical Devices recalls

All 13,884