CLASS II ONGOING Device recall · Reported 24 Sep 2025 · FDA Enforcement Report

The product is a handheld ultraviolet-C germicidal wand recalled

BigTree Sales is recalling The product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter. The recall was initiated on 29 Aug 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSKU BT-GERMWAND-LRG-WHT and Testing: September 8-10, 2020
Quantity recalledN/A

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2571-2025
FDA event ID
97621
Recalling firm
BigTree Sales, Arcadia, CA
Classification
Class II
Status
Ongoing
Recall initiated
29 Aug 2025
FDA classified
18 Sep 2025
Reported
24 Sep 2025
Type
FDA Mandated
Customers notified by
N/A
Reason category
Device Malfunction

Timeline

29 Aug 2025Recall initiated by company
18 Sep 2025Classified by FDA+20 days
24 Sep 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the accomplishment of the product s primary purpose of disinfecting objects and surfaces, and create a risk of skin and/or eye injuries to user and nearby persons.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

The product is a handheld ultraviolet-C germicidal wand and contains a narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.

Where it was distributed

U.S.A

Questions about this recall

Why was it recalled?

The subject product emits unsafe level of UVC radiation (i.e., the product reaches the exposure limit for Risk Group 3 (high risk) in about 2.2 minutes). The product has a radiation safety defect, in that these radiation emissions are unnecessary to the accomplishment of the product s primary purpose of disinfecting objects and surfaces, and create a risk of skin and/or eye injuries to user and nearby persons.

Which lots are affected?

SKU BT-GERMWAND-LRG-WHT and Testing: September 8-10, 2020

Where was it sold?

U.S.A

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.