CLASS II Device recall · Reported 25 Jul 2018 · FDA Enforcement Report

Proteus 235, Universal Nozzle with snout recalled

Ion Beam Applications is recalling Proteus 235, Universal Nozzle with snout ; PBS dedicated nozzle with snout XL, distributed nationwide in the US. The recall was initiated on 27 Jun 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.110, PAT.111, PAT.113, PAT.114, PAT.115, PAT.116, PAT.120, PAT.126, and PAT.132.
Quantity recalled14

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2470-2018
FDA event ID
80384
Recalling firm
Ion Beam Applications, Louvain-la-neuve
Classification
Class II
Status
Terminated
Recall initiated
27 Jun 2018
FDA classified
17 Jul 2018
Reported
25 Jul 2018
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Other Reasons

Timeline

27 Jun 2018Recall initiated by company
17 Jul 2018Classified by FDA+20 days
25 Jul 2018Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Gantry rotation or proton irradiation can be allowed even if apertures or accessories are not correctly positioned in snout 300x400 or snout XL. This can lead to risk of crushing for patient and/or user, and risk of mistreatment for the patient.

Product as listed by the FDA

Proteus 235, Universal Nozzle with snout 300x400; PBS dedicated nozzle with snout XL Product Usage: Device is designed to produce and deliver proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.

Where it was distributed

US Nationwide Distribution in the states of FL, VA, PA, OK, NJ, WA, and TX.

Nationwide FloridaNew JerseyOklahomaPennsylvaniaTexasVirginiaWashington

Questions about this recall

Why was it recalled?

Gantry rotation or proton irradiation can be allowed even if apertures or accessories are not correctly positioned in snout 300x400 or snout XL. This can lead to risk of crushing for patient and/or user, and risk of mistreatment for the patient.

Which lots are affected?

PAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.110, PAT.111, PAT.113, PAT.114, PAT.115, PAT.116, PAT.120, PAT.126, and PAT.132.

Where was it sold?

US Nationwide Distribution in the states of FL, VA, PA, OK, NJ, WA, and TX.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.