CLASS I ONGOING Device recall · Reported 14 Aug 2024 · FDA Enforcement Report

Puritan Bennett 500 Series Ventilators (Description recalled by Covidien

Covidien is recalling Puritan Bennett 500 Series Ventilators (Description, distributed in 34 states. The recall was initiated on 24 Jun 2024 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesREF/UDI-DI(GTIN): 4096600/10884521087798 4096600-01/10884521195998 4096600-02/10884521196001 4096600-03/10884521196018 4096600-04/10884521196025 4096600-05/10884521196032 4096600-06/10884521196377 4097100/4097100 4098300/10884521183636 4098300-01/10884521195943 4098300-02/4098300-02 4098300-03/10884521195967 4098300-05/10884521195981 4098300-06/10884521196384 DL4096600/10884521786509 DL4096600-06/10884521786516 DL4097100/DL4097100 DS4097100/DS4097100 All device serial numbers used before distribution of PB500 Series Ventilator User Manual and Clinician's Manual addendum, PT00196729 A, 2024-06-11
Quantity recalled34,642

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2446-2024
FDA event ID
94855
Recalling firm
Covidien, Carlsbad, CA
Classification
Class I
Status
Ongoing
Recall initiated
24 Jun 2024
FDA classified
2 Aug 2024
Reported
14 Aug 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

24 Jun 2024Recall initiated by company
2 Aug 2024Classified by FDA+39 days
14 Aug 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway

Product as listed by the FDA

Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600, PURITAN BENNETT PB560 VENTILATOR EU-DIV/4096600-01, PURITAN BENNETT 560 VENTILATOR JAPAN/4096600-02, PURITAN BENNETT 560 VENTILATOR APAC/4096600-03, Puritan Bennett 560 Ventilator LATAM/4096600-04, Puritan Bennett 560 Ventilator EMEA/4096600-05, Puritan Bennett 560 Ventilator CAN ANZ/4096600-06, Puritan Bennett 540 Ventilator USA/4097100 Puritan Bennett 520 Ventilator/4098300, Puritan Bennett 520 Ventilator EU-DIV/4098300-01, Puritan Bennett 520 Ventilator Japan/4098300-02, Puritan Bennett 520 Ventilator APAC/4098300-03, Puritan Bennett 520 Ventilator EMEA/4098300-05, Puritan Bennett 520 Ventilator CAN/ANZ/4098300-06, DL4096600 DL PB560 VENTILATOR/DL4096600, DL4096600-06 RFB PB560 VENT CAN ANZ/DL4096600-06, REFURBISHED PB540 VENTILATOR/DL4097100, Puritan Bennett 540 Ventilator Refurbished/DS4097100

Where it was distributed

US: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR, TX, WI, PA, CO, MI, NM, RI, VA, DC, VT, NC, NH, WV, NV, IA. OUS: Argentina, Australia, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Lithuania, Malaysia, Mexico, Montenegro, Namibia, Nepal, New Zealand, Nigeria, Panama, Poland, Reunion, Romania, Russian Federation, Saudi Arabia, Slovakia, Slovenia, Spain, Sweden, Taiwan, Turkey, Ukraine, United Kingdom, Viet Nam, South Africa, Austria, Kenya, Latvia, Lebanon, Mongolia, Nicaragua, Thailand, Tunisia, Algeria, Jordan, Libya, Malawi, Qatar, Pakistan, United Arab Emirates, Oman, Peru, Philippines, Bhutan, Iran - Islamic Republic Of, Bolivia, Kuwait, Egypt, Switzerland, Sudan, Norway, Korea - Republic Of, Bahrain, Ethiopia, Paraguay

ArkansasCaliforniaColoradoDistrict of ColumbiaGeorgiaIowaIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesotaMissouriNorth CarolinaNew HampshireNew JerseyNew Mexico

Questions about this recall

Why was it recalled?

Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway

Which lots are affected?

REF/UDI-DI(GTIN): 4096600/10884521087798 4096600-01/10884521195998 4096600-02/10884521196001 4096600-03/10884521196018 4096600-04/10884521196025 4096600-05/10884521196032 4096600-06/10884521196377 4097100/4097100 4098300/10884521183636 4098300-01/10884521195943 4098300-02/4098300-02 4098300-03/10884521195967 4098300-05/10884521195981 4098300-06/10884521196384 DL4096600/10884521786509 DL4096600-06/10884521786516 DL4097100/DL4097100 DS4097100/DS4097100 All device serial numbers used before distribution of PB500 Series Ventilator User Manual and Clinician's Manual addendum, PT00196729 A, 2024-06-11

Where was it sold?

US: KY, OH, GA, MN, IN,UT, IL, NJ, NY, PR, LA, MD, MO, TN, MA, ME, CA, SC, AR, TX, WI, PA, CO, MI, NM, RI, VA, DC, VT, NC, NH, WV, NV, IA. OUS: Argentina, Australia, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Lithuania, Malaysia, Mexico, Montenegro, Namibia, Nepal, New Zealand, Nigeria, Panama, Poland, Reunion, Romania, Russian Federation, Saudi Arabia, Slovakia, Slovenia, Spain, Sweden, Taiwan, Turkey, Ukraine, United Kingdom, Viet Nam, South Africa, Austria, Kenya, Latvia, Lebanon, Mongolia, Nicaragua, Thailand, Tunisia, Algeria, Jordan, Libya, Malawi, Qatar, Pakistan, United Arab Emirates, Oman, Peru, Philippines, Bhutan, Iran - Islamic Republic Of, Bolivia, Kuwait, Egypt, Switzerland, Sudan, Norway, Korea - Republic Of, Bahrain, Ethiopia, Paraguay

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Covidien

All 493

Other CPAP & Ventilators recalls

All 554