CLASS II Device recall · Reported 2 Jul 2014 · FDA Enforcement Report
Puritan Bennett 840 Ventilator recalled
Nellcor Puritan Bennett Inc. ( is recalling Puritan Bennett 840 Ventilator (4-840120DIUU-XX) The PB840 Ventilator System is intended, distributed nationwide in the US. The recall was initiated on 3 Jun 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1844-2014
- FDA event ID
- 68444
- Recalling firm
- Nellcor Puritan Bennett Inc. (, Boulder, CO
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 3 Jun 2014
- FDA classified
- 20 Jun 2014
- Reported
- 2 Jul 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- CPAP & Ventilators
Timeline
| 3 Jun 2014 | Recall initiated by company | |
| 20 Jun 2014 | Classified by FDA | +17 days |
| 2 Jul 2014 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Covidien is conducting a medical device field correction for specific Puritan Bennett 840 Ventilators due to customer reports of loss of graphical user interface (GUI) display information while the ventilator continues to provide breath support.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Puritan Bennett 840 Ventilator (4-840120DIUU-XX) The PB840 Ventilator System is intended to provide invasive or non-invasive ventilatory support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency. It is used in hospitals and healthcare facilities.
Where it was distributed
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY., and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Egypt, Germany, Guatemala, Israel, Japan, Kazakhstan, Kenya, Lebanon, Libya, Macedonia, Mexico, Montenegro, Morocco, Panama, Peru, Russia, Saudi Arabia, Singapore, Switzerland, Taiwan, Tunisia, Turkey, Yugoslavia (Serbia & Montenegro) Including GOVT/VA/Military
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichigan
Questions about this recall
Why was it recalled?
Covidien is conducting a medical device field correction for specific Puritan Bennett 840 Ventilators due to customer reports of loss of graphical user interface (GUI) display information while the ventilator continues to provide breath support.
Which lots are affected?
serial number 3510100714 to 3512101160 and any units in which a replacement PSU was installed between March 2010 and September of 2010.
Where was it sold?
Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY., and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, Costa Rica, Egypt, Germany, Guatemala, Israel, Japan, Kazakhstan, Kenya, Lebanon, Libya, Macedonia, Mexico, Montenegro, Morocco, Panama, Peru, Russia, Saudi Arabia, Singapore, Switzerland, Taiwan, Tunisia, Turkey, Yugoslavia (Serbia & Montenegro) Including GOVT/VA/Military
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Nellcor Puritan Bennett Inc. (
All 11| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Covidien Puritan Bennett 980 Ventilator recalled | Nellcor Puritan Bennett Inc. ( | Other | Nationwide |
| Device | CLASS I | Covidien Puritan Bennett 980 Ventilator recalled | Nellcor Puritan Bennett Inc. ( | Device Malfunction | 17 states |
| Device | CLASS II | Vital Sync RMS 1.0 when loaded with v2.4 recalled | Nellcor Puritan Bennett Inc. ( | Device Malfunction | 8 states |
| Device | CLASS II | Vital Sync" VPMP & IM recalled by Nellcor Puritan Bennett Inc. ( | Nellcor Puritan Bennett Inc. ( | Device Malfunction | 8 states |
| Device | CLASS II | Model Number(s): PB840 Ventilator recalled | Nellcor Puritan Bennett Inc. ( | Device Malfunction | Nationwide |
Other CPAP & Ventilators recalls
All 552| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Getinge Y Sensor Module, Servo-u, Servo recalled | Maquet Critical Care AB | Device Malfunction | Nationwide |
| Device | CLASS II | Biofire Respiratory Panel 2.1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |
| Device | CLASS I | Astral 100/150 Resmed Astral 100/50 is a ventilator system recalled | ResMed | Packaging Defects | Nationwide |
| Device | CLASS I | Breathing circuit set, coaxial recalled by Hamilton Medical AG | Hamilton Medical | Device Malfunction | Nationwide |