CLASS II ONGOING Device recall · Reported 21 Sep 2022 · FDA Enforcement Report
QIAcube Connect MDx recalled by Qiagen Sciences
Qiagen Sciences is recalling QIAcube Connect MDx, distributed nationwide in the US. The recall was initiated on 21 Nov 2021 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1746-2022
- FDA event ID
- 90758
- Recalling firm
- Qiagen Sciences, Germantown, MD
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 21 Nov 2021
- FDA classified
- 15 Sep 2022
- Reported
- 21 Sep 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Lead & Heavy Metals
- Device type
- Other Medical Devices
Timeline
| 21 Nov 2021 | Recall initiated by company | |
| 15 Sep 2022 | Classified by FDA | +298 days |
| 21 Sep 2022 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
QIAcube Connect MDx, Model No. 9003070
Where it was distributed
Domestic distribution to AZ, CA, CO, DE, DC, FL, GA, IA, KS, LA, MA, MD, MN, NV, NJ, NY, NC, OH, PA, SC, SD, TN, TX, VA, WA, WI. International distribution worldwide.
Nationwide ArizonaCaliforniaColoradoDistrict of ColumbiaDelawareFloridaGeorgiaIowaKansasLouisianaMassachusettsMarylandMinnesotaNorth CarolinaNew JerseyNevadaNew YorkOhioPennsylvaniaSouth Carolina
Questions about this recall
Is QIAcube Connect MDx recalled?
Yes. Qiagen Sciences recalled QIAcube Connect MDx on 21 Nov 2021. The recall is Class II and its status is Ongoing. Recall number Z-1746-2022.
Why was it recalled?
During the "Load tip racks and enzymes" step of the run set-up, the info screen indicates "minimal volume to be loaded", but the indicated volume is instead the exact volume that should be loaded. Misunderstanding the guidance may lead to over diluted, under diluted, or improperly lysed sample, which in turn could lead to false negative or false positive results.
Which lots are affected?
UDI-DI (GTIN): 04053228039129 Serial Numbers (U.S.) 30000 30009 30027 30039 30049 30061 30073 30002 30010 30028 30041 30050 30062 30075 30003 30132 30031 30042 30051 30063 30078 30004 30013 30032 30043 30053 30064 30080 30005 30016 30034 30044 30057 30065 30082 30006 30019 30035 30045 30058 30067 30089 30007 30021 30033 30046 30059 30069 30090 30008 30023 30037 30047 30060 30071 30093 30094 30095 30097 30102 30104 30109 30112 30114 30115 30117 30118 30119 30121 30124 30127 30128 30129 30130 30131 30133 30134 30136 30138 30139 30140 30141 30143 30145 30106 30025 30077 30054 30055 30066 30076
Where was it sold?
Domestic distribution to AZ, CA, CO, DE, DC, FL, GA, IA, KS, LA, MA, MD, MN, NV, NJ, NY, NC, OH, PA, SC, SD, TN, TX, VA, WA, WI. International distribution worldwide.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Qiagen Sciences
All 23| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | QIAstat-Dx Respiratory Panel Plus Reference Number recalled | Qiagen Sciences | Device Malfunction | 12 states |
| Device | CLASS II | QIAcube Connect MDx - IVD Instrument designed to perform recalled | Qiagen Sciences | Potency & Assay | Nationwide |
| Device | CLASS II | EZ2 Connect MDx-IVD Designed to perform automated isolation recalled | Qiagen Sciences | Temperature Abuse | Nationwide |
| Device | CLASS II | QIAstat-Dx Respiratory Sars-CoV-2 Panel multiplexed nucleic recalled | Qiagen Sciences | Other | Nationwide |
| Device | CLASS II | therascreen Fgfr Rgq Rt-pcr kit (US Ivd) recalled by Qiagen Sciences | Qiagen Sciences | Lead & Heavy Metals | 4 states |
Other Other Medical Devices recalls
All 13,771| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: Blood Draw Kit recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ortho Aspiration Kit recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep DYKA1054F recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Arterial Line Insertion Kit recalled | Medline Industries | Sterility | Nationwide |