CLASS II Device recall · Reported 7 Nov 2012 · FDA Enforcement Report

Quadrox iD Hmod 2030 Diffusion Membrane Oxygenator recalled

Maquet Cardiovascular Us Sales is recalling Quadrox iD Hmod 2030 Diffusion Membrane Oxygenator, distributed nationwide in the US. The recall was initiated on 12 Jun 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll sizes and membrane types of QUADROX-i oxygenators, i.e., QUADROX-i(D) Adult, Small Adult, Pediatric and Neonatal with SOFTLINE and BIOLINE Coating; Catalogue numbers to include BE-HMOD XXXXX; BEQ-HMO XXXXX; BEQ-HMOD XXXX; BEQ-HMOD XXXXX; BO-VKMO XXXXXX; HMO XXXX; HMO XXXXX; HMOD XXXX; HMOD XXXXX; VKMO XXXX; VKMO XXXXX; X HMO XXXXX U; XHMOD 30000 U; X VKMO XXXXX U; and X 1895. QUADROX-iR with SOFTLINE and BIOLINE Coating - Catalogue numbers to include XVIVO HMO 70100 U. HLS module 5.0 and 7.0 in HLS Sets: Catalogue numbers to include BEQ-HLS 5050 and BEQ-HLS 7050. HLS module 5.0 and 7.0 in HIT HLS Sets: Catalogue numbers to include BO-HLS 5050 and BO-HLS 7050. Custom Tubing Packs containing oxygenators listed above: Catalogue numbers to include TOP XXXXX; TOP XXXXXns; TOP XXXX; BO-TOP XXXXX; BEQ TOP XXXXX; BSQ-TOP XXXX; and BSQ-TOP XXXXX. Note: an X was used to represent variations of catalogue numbers.
Quantity recalled23,203 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0160-2013
FDA event ID
62947
Recalling firm
Maquet Cardiovascular Us Sales, Wayne, NJ
Classification
Class II
Status
Terminated
Recall initiated
12 Jun 2012
FDA classified
26 Oct 2012
Reported
7 Nov 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Temperature Abuse

Timeline

12 Jun 2012Recall initiated by company
26 Oct 2012Classified by FDA+136 days
7 Nov 2012Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped.

Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.

Product as listed by the FDA

QUADROX iD HMOD 2030 Diffusion Membrane Oxygenator; QUADROX-i ADULT HMO 70000 / HMO 71000; QUADROX-iD Pediatric HMOD 30000 Product Usage: The QUADROX oxygenators are intended for the use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, eliminates carbon dioxide and regulates blood temperature. The devices utilization period is limited to six hours.

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Is Quadrox iD Hmod 2030 Diffusion Membrane Oxygenator recalled?

Yes. Maquet Cardiovascular Us Sales recalled Quadrox iD Hmod 2030 Diffusion Membrane Oxygenator on 12 Jun 2012. The recall is Class II and its status is Terminated. Recall number Z-0160-2013.

Why was it recalled?

It has come to firm's attention that, in some rare cases, the blood outlet connector on some models of MAQUET Oxygenators may detach from the oxygenator. This event has occurred in about 0.01% (1:10,000) of units shipped.

Which lots are affected?

All sizes and membrane types of QUADROX-i oxygenators, i.e., QUADROX-i(D) Adult, Small Adult, Pediatric and Neonatal with SOFTLINE and BIOLINE Coating; Catalogue numbers to include BE-HMOD XXXXX; BEQ-HMO XXXXX; BEQ-HMOD XXXX; BEQ-HMOD XXXXX; BO-VKMO XXXXXX; HMO XXXX; HMO XXXXX; HMOD XXXX; HMOD XXXXX; VKMO XXXX; VKMO XXXXX; X HMO XXXXX U; XHMOD 30000 U; X VKMO XXXXX U; and X 1895. QUADROX-iR with SOFTLINE and BIOLINE Coating - Catalogue numbers to include XVIVO HMO 70100 U. HLS module 5.0 and 7.0 in HLS Sets: Catalogue numbers to include BEQ-HLS 5050 and BEQ-HLS 7050. HLS module 5.0 and 7.0 in HIT HLS Sets: Catalogue numbers to include BO-HLS 5050 and BO-HLS 7050. Custom Tubing Packs containing oxygenators listed above: Catalogue numbers to include TOP XXXXX; TOP XXXXXns; TOP XXXX; BO-TOP XXXXX; BEQ TOP XXXXX; BSQ-TOP XXXX; and BSQ-TOP XXXXX. Note: an X was used to represent variations of catalogue numbers.

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Maquet Cardiovascular Us Sales

All 62

Other Other Medical Devices recalls

All 13,778