CLASS II ONGOING Device recall · Reported 8 Apr 2026 · FDA Enforcement Report

RefleXion X1 Radiotherapy System recalled by Reflexion Medical

Reflexion Medical is recalling RefleXion X1 Radiotherapy System, distributed in 8 states. The recall was initiated on 9 Feb 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.
Quantity recalled12

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1656-2026
FDA event ID
98518
Recalling firm
Reflexion Medical, Hayward, CA
Classification
Class II
Status
Ongoing
Recall initiated
9 Feb 2026
FDA classified
27 Mar 2026
Reported
8 Apr 2026
Type
Voluntary: Firm initiated
Reason category
Device Malfunction

Timeline

9 Feb 2026Recall initiated by company
27 Mar 2026Classified by FDA+46 days
8 Apr 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

Where it was distributed

US: CA, TX, PA, CT, NJ, OR, LA, OH

CaliforniaConnecticutLouisianaNew JerseyOhioOregonPennsylvaniaTexas

Questions about this recall

Why was it recalled?

Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.

Which lots are affected?

Software Version: 2.1.29-4 and Software Version: 2.1.35-1. UDI: (01)00860003983812(11)230504(21)X11001, (01)00860003983812(11)230518(21)X11002, (01)00860003983812(11)230505(21)X11003, (01)00860003983812(11)230530(21)X11004, (01)00860003983812(11)230811(21)X11006, (01)00860003983812(11)230410(21)X11007, (01)00860003983812(11)230803(21)X11008, (01)00860003983812(11)240528(21)X11009, (01)00860003983812(11)240820(21)X11010, (01)00860003983812(11)241107(21)X11011, (01)00860003983812(11)241206(21)X11012, (01)00860003983812(11)250731(21)X11013.

Where was it sold?

US: CA, TX, PA, CT, NJ, OR, LA, OH

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Reflexion Medical

All 4

Other Other Medical Devices recalls

All 13,884