CLASS I ONGOING Device recall · Reported 12 Feb 2025 · FDA Enforcement Report

regard Item Number: 830106006, LD01114F - Newborn Kit containing recalled

ROi CPS is recalling regard Item Number: 830106006, LD01114F - Newborn Kit containing the, distributed in Missouri. The recall was initiated on 26 Dec 2024 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesLot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.
Quantity recalled414 kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1047-2025
FDA event ID
96094
Recalling firm
ROi CPS, Republic, MO
Classification
Class I
Status
Ongoing
Recall initiated
26 Dec 2024
FDA classified
4 Feb 2025
Reported
12 Feb 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

26 Dec 2024Recall initiated by company
4 Feb 2025Classified by FDA+40 days
12 Feb 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

regard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.

Where it was distributed

Distribution was made to MO. There was no government/military/foreign distribution.

Missouri

Questions about this recall

Why was it recalled?

There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..

Which lots are affected?

Lot number 104006, Exp. 9/30/2026, UDI (01)10194717119074(17)260930(10)104006; and Lot number 103486, Exp. 8/31/2026, UDI (01)10194717119074(17)260831(10)103486.

Where was it sold?

Distribution was made to MO. There was no government/military/foreign distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from ROi CPS

All 86

Other Other Medical Devices recalls

All 13,884