CLASS I ONGOING Device recall · Reported 8 Oct 2025 · FDA Enforcement Report
Reprocessed Livewire Steerable Electrophysiology Catheter recalled
Medline Industries is recalling Reprocessed Livewire Steerable Electrophysiology Catheter, distributed nationwide in the US. The recall was initiated on 31 Jul 2025 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2610-2025
- FDA event ID
- 97496
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 31 Jul 2025
- FDA classified
- 26 Sep 2025
- Reported
- 8 Oct 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Particulate Matter
- Device type
- Heart & Vascular Devices
Timeline
| 31 Jul 2025 | Recall initiated by company | |
| 26 Sep 2025 | Classified by FDA | +57 days |
| 8 Oct 2025 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Product as listed by the FDA
Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH
Where it was distributed
Domestic US distribution nationwide. International distribution pending. No international distribution.
Questions about this recall
Why was it recalled?
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Which lots are affected?
Item Number/UDI (ea)/UDI(case): 401575RH 10888277407497 20888277407494; 401582RH 10888277407510 20888277407517; 401600RH 10888277407541 20888277407548; 401603RH 10888277407558 20888277407555; 401606RH 10888277407565 20888277407562; 401652RH 10888277407572 20888277407579; 401653RH 10888277407589 20888277407586; 401654RH 10888277407596 20888277407593; 401904RH 10888277407602 20888277407609; 401905RH 10888277407619 20888277407616; 401908RH 10197344020966 20197344020963; 401914RH 10888277407527 20888177407524; 401915RH 10888277407626 20888277407623; 401918RH 10888277407640 20888277407647; 401932RH 10888277407671 20888277407678; 401933RH 10888277407688 20888277407685; 401934RH 10888277407695 20888277407692; 401938RH 10888277407701 20888277407708; 401939RH 10888277407718 20888277407715; 401940RH 10888277407480 20888277407487; 401941RH 10888277407725 20888277407722; 401991RH 10888277407756 2088277407753; LOTS EP230724 EP230731 EP230803 EP230816 EP230914 EP231017 EP231116 EP240108 EP240122 EP240212 EP240227 EP240311 EP240322 EP240411 EP240415 EP240422 EP240429 EP240507 EP240514 EP240520 EP240528 EP240607 EP240611 EP240710 EP240715 EP240723 EP240805 EP240812 EP240820 EP240829 EP240911 EP240923 EP240930 EP241008 EP241030 EP241125 EP250108
Where was it sold?
Domestic US distribution nationwide. International distribution pending. No international distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
4 other products · FDA event 97496| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS I | Reprocessed Webster CS Catheter 10 Electrodes recalled | Medline Industries | Particulate Matter | Nationwide | |
| CLASS I | Reprocessed Inquiry Steerable Diagnostics Catheter recalled | Medline Industries | Particulate Matter | Nationwide | |
| CLASS I | Reprocessed Acuson AcuNav Ultrasound Catheter recalled | Medline Industries | Particulate Matter | Nationwide | |
| CLASS I | Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac recalled | Medline Industries | Particulate Matter | Nationwide |
More recalls from Medline Industries
All 738| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Drape C-section W/Pouch ST 7/CS, Medline Product Number recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: Fetal Surgery Cds-lf recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Drape C-section Fen Clr Screen W, Medline Product Number recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: C-section Pack recalled | Medline Industries | Other | Nationwide |
Other Heart & Vascular Devices recalls
All 6,271| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Cardinal Health Presource PBDS kits recalled | Cardinal Health 200 | Other | IL |
| Device | CLASS I | Impella Connect Optical AIC, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Impella Optical Controller, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Automated Impella Controller, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS II | Direxion Torqueable Microcatheter, Ngmc/Fathom/027/Straight/1RO/130 recalled | Boston Scientific Cork | Device Malfunction | Nationwide |