CLASS I ONGOING Device recall · Reported 8 Oct 2025 · FDA Enforcement Report

Reprocessed Livewire Steerable Electrophysiology Catheter recalled

Medline Industries is recalling Reprocessed Livewire Steerable Electrophysiology Catheter, distributed nationwide in the US. The recall was initiated on 31 Jul 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesItem Number/UDI (ea)/UDI(case): 401575RH 10888277407497 20888277407494; 401582RH 10888277407510 20888277407517; 401600RH 10888277407541 20888277407548; 401603RH 10888277407558 20888277407555; 401606RH 10888277407565 20888277407562; 401652RH 10888277407572 20888277407579; 401653RH 10888277407589 20888277407586; 401654RH 10888277407596 20888277407593; 401904RH 10888277407602 20888277407609; 401905RH 10888277407619 20888277407616; 401908RH 10197344020966 20197344020963; 401914RH 10888277407527 20888177407524; 401915RH 10888277407626 20888277407623; 401918RH 10888277407640 20888277407647; 401932RH 10888277407671 20888277407678; 401933RH 10888277407688 20888277407685; 401934RH 10888277407695 20888277407692; 401938RH 10888277407701 20888277407708; 401939RH 10888277407718 20888277407715; 401940RH 10888277407480 20888277407487; 401941RH 10888277407725 20888277407722; 401991RH 10888277407756 2088277407753; LOTS EP230724 EP230731 EP230803 EP230816 EP230914 EP231017 EP231116 EP240108 EP240122 EP240212 EP240227 EP240311 EP240322 EP240411 EP240415 EP240422 EP240429 EP240507 EP240514 EP240520 EP240528 EP240607 EP240611 EP240710 EP240715 EP240723 EP240805 EP240812 EP240820 EP240829 EP240911 EP240923 EP240930 EP241008 EP241030 EP241125 EP250108
Quantity recalled7913 eaches

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2610-2025
FDA event ID
97496
Recalling firm
Medline Industries, Northfield, IL
Classification
Class I
Status
Ongoing
Recall initiated
31 Jul 2025
FDA classified
26 Sep 2025
Reported
8 Oct 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Particulate Matter

Timeline

31 Jul 2025Recall initiated by company
26 Sep 2025Classified by FDA+57 days
8 Oct 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Product as listed by the FDA

Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH

Where it was distributed

Domestic US distribution nationwide. International distribution pending. No international distribution.

Nationwide

Questions about this recall

Why was it recalled?

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Which lots are affected?

Item Number/UDI (ea)/UDI(case): 401575RH 10888277407497 20888277407494; 401582RH 10888277407510 20888277407517; 401600RH 10888277407541 20888277407548; 401603RH 10888277407558 20888277407555; 401606RH 10888277407565 20888277407562; 401652RH 10888277407572 20888277407579; 401653RH 10888277407589 20888277407586; 401654RH 10888277407596 20888277407593; 401904RH 10888277407602 20888277407609; 401905RH 10888277407619 20888277407616; 401908RH 10197344020966 20197344020963; 401914RH 10888277407527 20888177407524; 401915RH 10888277407626 20888277407623; 401918RH 10888277407640 20888277407647; 401932RH 10888277407671 20888277407678; 401933RH 10888277407688 20888277407685; 401934RH 10888277407695 20888277407692; 401938RH 10888277407701 20888277407708; 401939RH 10888277407718 20888277407715; 401940RH 10888277407480 20888277407487; 401941RH 10888277407725 20888277407722; 401991RH 10888277407756 2088277407753; LOTS EP230724 EP230731 EP230803 EP230816 EP230914 EP231017 EP231116 EP240108 EP240122 EP240212 EP240227 EP240311 EP240322 EP240411 EP240415 EP240422 EP240429 EP240507 EP240514 EP240520 EP240528 EP240607 EP240611 EP240710 EP240715 EP240723 EP240805 EP240812 EP240820 EP240829 EP240911 EP240923 EP240930 EP241008 EP241030 EP241125 EP250108

Where was it sold?

Domestic US distribution nationwide. International distribution pending. No international distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

4 other products · FDA event 97496

More recalls from Medline Industries

All 738

Other Heart & Vascular Devices recalls

All 6,271