CLASS I ONGOING Device recall · Reported 8 Oct 2025 · FDA Enforcement Report
Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac recalled
Medline Industries is recalling Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures, distributed nationwide in the US. The recall was initiated on 31 Jul 2025 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2613-2025
- FDA event ID
- 97496
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 31 Jul 2025
- FDA classified
- 26 Sep 2025
- Reported
- 8 Oct 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Particulate Matter
- Device type
- Heart & Vascular Devices
Timeline
| 31 Jul 2025 | Recall initiated by company | |
| 26 Sep 2025 | Classified by FDA | +57 days |
| 8 Oct 2025 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Product as listed by the FDA
Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH
Where it was distributed
Domestic US distribution nationwide. International distribution pending. No international distribution.
Questions about this recall
Why was it recalled?
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Which lots are affected?
Item Number/UDI-DI D087031RH 10197344019557 LOTS EP240731 EP240814 EP240826 EP240903 EP240916 EP241003 EP241014 EP241028 EP241104 EP241112 EP241202 EP241209A EP241219 EP241224 EP250114 EP250116 EP250204 EP250219
Where was it sold?
Domestic US distribution nationwide. International distribution pending. No international distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
4 other products · FDA event 97496This product is part of a recall event; the main page for the event is Reprocessed Livewire Steerable Electrophysiology Catheter recalled.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS I | Reprocessed Livewire Steerable Electrophysiology Catheter recalled | Medline Industries | Particulate Matter | Nationwide | |
| CLASS I | Reprocessed Webster CS Catheter 10 Electrodes recalled | Medline Industries | Particulate Matter | Nationwide | |
| CLASS I | Reprocessed Inquiry Steerable Diagnostics Catheter recalled | Medline Industries | Particulate Matter | Nationwide | |
| CLASS I | Reprocessed Acuson AcuNav Ultrasound Catheter recalled | Medline Industries | Particulate Matter | Nationwide |
More recalls from Medline Industries
All 738| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Drape C-section W/Pouch ST 7/CS, Medline Product Number recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: Fetal Surgery Cds-lf recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Drape C-section Fen Clr Screen W, Medline Product Number recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline medical convenience kits labeled as: C-section Pack recalled | Medline Industries | Other | Nationwide |
Other Heart & Vascular Devices recalls
All 6,271| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Cardinal Health Presource PBDS kits recalled | Cardinal Health 200 | Other | IL |
| Device | CLASS I | Impella Connect Optical AIC, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Impella Optical Controller, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS I | Automated Impella Controller, Packaged recalled by Abiomed | Abiomed | Potency & Assay | Nationwide |
| Device | CLASS II | Direxion Torqueable Microcatheter, Ngmc/Fathom/027/Straight/1RO/130 recalled | Boston Scientific Cork | Device Malfunction | Nationwide |