CLASS III ONGOING Device recall · Reported 21 Aug 2024 · FDA Enforcement Report

Salivary Testosterone Elisa recalled by DRG International

DRG International is recalling Salivary Testosterone Elisa, distributed nationwide in the US. The recall was initiated on 5 Jul 2024 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesModel/Catalog Number: SLV-3013; UDI-DI: 04048474030136; Lot Nymbers: 66K014 and 66K123.
Quantity recalled257 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2622-2024
FDA event ID
95021
Recalling firm
DRG International, Springfield, NJ
Classification
Class III
Status
Ongoing
Recall initiated
5 Jul 2024
FDA classified
15 Aug 2024
Reported
21 Aug 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

5 Jul 2024Recall initiated by company
15 Aug 2024Classified by FDA+41 days
21 Aug 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Invalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Product Name: Salivary Testosterone ELISA Model/Catalog Number: SLV-3013 Product Description: An enzyme immunoassay for the quantitative in vitro diagnostic measurement of free active testosterone in saliva.

Where it was distributed

US Nationwide distribution in the states of CA, IL, MN, PA, UT, VA.

Nationwide CaliforniaIllinoisMinnesotaPennsylvaniaUtahVirginia

Questions about this recall

Why was it recalled?

Invalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients.

Which lots are affected?

Model/Catalog Number: SLV-3013; UDI-DI: 04048474030136; Lot Nymbers: 66K014 and 66K123.

Where was it sold?

US Nationwide distribution in the states of CA, IL, MN, PA, UT, VA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from DRG International

All 7

Other Test Kits & Diagnostics recalls

All 2,663