CLASS II ONGOING Device recall · Reported 13 Nov 2024 · FDA Enforcement Report

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting recalled

Zimmer is recalling Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally, distributed nationwide in the US. The recall was initiated on 17 Sep 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: +M718VPX40001/$$326051543DD6738Q, +M718VPX4001/$$326062143DD8092J Lot Numbers: 43DD6738, 43DD8092
Quantity recalled432 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0285-2025
FDA event ID
95472
Recalling firm
Zimmer, Warsaw, IN
Classification
Class II
Status
Ongoing
Recall initiated
17 Sep 2024
FDA classified
1 Nov 2024
Reported
13 Nov 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

17 Sep 2024Recall initiated by company
1 Nov 2024Classified by FDA+45 days
13 Nov 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273)

Where it was distributed

Nationwide

Nationwide

Questions about this recall

Why was it recalled?

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery

Which lots are affected?

UDI: +M718VPX40001/$$326051543DD6738Q, +M718VPX4001/$$326062143DD8092J Lot Numbers: 43DD6738, 43DD8092

Where was it sold?

Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zimmer

All 539

Other Surgical Instruments recalls

All 3,891