CLASS II ONGOING Device recall · Reported 12 Feb 2025 · FDA Enforcement Report

SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part recalled

LivaNova USA is recalling SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the, distributed in 41 states. The recall was initiated on 12 Dec 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI-DI: 1000/05425025750405, 1000-D/05425025750528 All Serial Numbers less than 500,000
Quantity recalled81,606

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1072-2025
FDA event ID
95435
Recalling firm
LivaNova USA, Houston, TX
Classification
Class II
Status
Ongoing
Recall initiated
12 Dec 2024
FDA classified
31 Jan 2025
Reported
12 Feb 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

12 Dec 2024Recall initiated by company
31 Jan 2025Classified by FDA+50 days
12 Feb 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.

Product as listed by the FDA

SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).

Where it was distributed

US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.

AlaskaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaineMichiganMinnesotaMissouri

Questions about this recall

Is SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the recalled?

Yes. LivaNova USA recalled SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the on 12 Dec 2024. The recall is Class II and its status is Ongoing. Recall number Z-1072-2025.

Why was it recalled?

Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.

Which lots are affected?

REF/UDI-DI: 1000/05425025750405, 1000-D/05425025750528 All Serial Numbers less than 500,000

Where was it sold?

US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.