CLASS II ONGOING Device recall · Reported 11 Sep 2024 · FDA Enforcement Report

Sevoflurane E_Vap.(Safety Filling adapter) used recalled

Mindray DS USA is recalling Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to, distributed nationwide in the US. The recall was initiated on 22 Jul 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 06936415975777 Serial Number: AH6A28000471 AH6A24000202 AH6A28000474 AH6A27000378 AH6A28000466 AH6A24000195 AH6A28000457 AH6A25000237 AH6A28000426 AH6A24000199 AH6A24000207
Quantity recalled96 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2950-2024
FDA event ID
95099
Recalling firm
Mindray DS USA, Mahwah, NJ
Classification
Class II
Status
Ongoing
Recall initiated
22 Jul 2024
FDA classified
4 Sep 2024
Reported
11 Sep 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

22 Jul 2024Recall initiated by company
4 Sep 2024Classified by FDA+44 days
11 Sep 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: 115-066756-00

Where it was distributed

US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

Nationwide CaliforniaGeorgiaKentuckyMississippiNew JerseyPuerto RicoTexasVirginiaWest Virginia

Questions about this recall

Is Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to recalled?

Yes. Mindray DS USA recalled Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to on 22 Jul 2024. The recall is Class II and its status is Ongoing. Recall number Z-2950-2024.

Why was it recalled?

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Which lots are affected?

UDI-DI: 06936415975777 Serial Number: AH6A28000471 AH6A24000202 AH6A28000474 AH6A27000378 AH6A28000466 AH6A24000195 AH6A28000457 AH6A25000237 AH6A28000426 AH6A24000199 AH6A24000207

Where was it sold?

US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Mindray DS USA

All 39

Other Other Medical Devices recalls

All 13,778