CLASS II ONGOING Device recall · Reported 8 Jul 2026 · FDA Enforcement Report

Short Monobloc Stem Broaches, Size recalled by Medacta Usa

Medacta Usa is recalling Short Monobloc Stem Broaches, Size, distributed nationwide in the US. The recall was initiated on 30 Apr 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF/UDI/Lot: 04.01.10.0970/07630542768910/2585086, 04.01.10.0971/07630542768927/2585087, 04.01.10.0972/07630542768934/2585088, 04.01.10.0973/07630542768941/2585089, 04.01.10.0974/07630542768958/2585090, 04.01.10.0975/07630542768965/2585091, 04.01.10.0976/07630542768972/2585092, 04.01.10.0977/07630542768989/2585093, 04.01.10.0978/07630542768996/2585094, 04.01.10.0979/07630542769009/2585095
Quantity recalled50

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2618-2026
FDA event ID
98944
Recalling firm
Medacta Usa, Memphis, TN
Classification
Class II
Status
Ongoing
Recall initiated
30 Apr 2026
FDA classified
29 Jun 2026
Reported
8 Jul 2026
Type
Voluntary: Firm initiated
Customers notified by
N/A
Reason category
Device Malfunction

Timeline

30 Apr 2026Recall initiated by company
29 Jun 2026Classified by FDA+60 days
8 Jul 2026Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System

Where it was distributed

US Nationwide distribution in the states of CO, AR, IN, PA, UT, FL.

Nationwide ArkansasColoradoFloridaIndianaPennsylvaniaUtah

Questions about this recall

Why was it recalled?

Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.

Which lots are affected?

REF/UDI/Lot: 04.01.10.0970/07630542768910/2585086, 04.01.10.0971/07630542768927/2585087, 04.01.10.0972/07630542768934/2585088, 04.01.10.0973/07630542768941/2585089, 04.01.10.0974/07630542768958/2585090, 04.01.10.0975/07630542768965/2585091, 04.01.10.0976/07630542768972/2585092, 04.01.10.0977/07630542768989/2585093, 04.01.10.0978/07630542768996/2585094, 04.01.10.0979/07630542769009/2585095

Where was it sold?

US Nationwide distribution in the states of CO, AR, IN, PA, UT, FL.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Medacta Usa

All 20

Other Other Medical Devices recalls

All 13,968