CLASS II Device recall · Reported 29 Oct 2014 · FDA Enforcement Report

Siemens Advia Centaur¿ Systems HBc Total (HBcT), Reagent recalled

Siemens Healthcare Diagnostics is recalling Siemens Advia Centaur¿ Systems HBc Total (HBcT), Reagent, distributed nationwide in the US. The recall was initiated on 15 Sep 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Number: 100064 Exp.Date:February 10, 2015
Quantity recalled5434 (200 tests)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0102-2015
FDA event ID
69292
Recalling firm
Siemens Healthcare Diagnostics, East Walpole, MA
Classification
Class II
Status
Terminated
Recall initiated
15 Sep 2014
FDA classified
21 Oct 2014
Reported
29 Oct 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Hepatitis A

Timeline

15 Sep 2014Recall initiated by company
21 Oct 2014Classified by FDA+36 days
29 Oct 2014Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Increased imprecision and positive bias is observed with patient samples and negative controls and can result in an increase in initial reactive results.

Product as listed by the FDA

Siemens ADVIA Centaur¿ Systems HBc Total (HBcT), Reagent. For In Vitro Diagnostic Use Catalog Number: 07566733 Siemens Material Number: 10309508 The ADVIA Centaur HBc Total assay is an in vitro diagnostic test for the qualitative determination of total antibodies to the core antigen of the hepatitis B virus (HBc Total) in human serum or plasma (potassium EDTA, or lithium or sodium heparinized) using the ADVIA Centaur and ADVIA Centaur XP system.

Where it was distributed

Worldwide Distribution-USA (nationwide) and the countries of Austria Bahrain Belgium Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iraq Ireland Italy Jordan Latvia Lebanon Lithuania Luxembourg Netherlands Norway Poland Portugal Romania Slovakia Spain Sweden Switzerland Turkey Unit.Arab Emir and Yemen.

Nationwide

Questions about this recall

Is Siemens Advia Centaur¿ Systems HBc Total (HBcT), Reagent recalled?

Yes. Siemens Healthcare Diagnostics recalled Siemens Advia Centaur¿ Systems HBc Total (HBcT), Reagent on 15 Sep 2014. The recall is Class II and its status is Terminated. Recall number Z-0102-2015.

Why was it recalled?

Increased imprecision and positive bias is observed with patient samples and negative controls and can result in an increase in initial reactive results.

Which lots are affected?

Lot Number: 100064 Exp.Date:February 10, 2015

Where was it sold?

Worldwide Distribution-USA (nationwide) and the countries of Austria Bahrain Belgium Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iraq Ireland Italy Jordan Latvia Lebanon Lithuania Luxembourg Netherlands Norway Poland Portugal Romania Slovakia Spain Sweden Switzerland Turkey Unit.Arab Emir and Yemen.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Siemens Healthcare Diagnostics

All 694

Other Test Kits & Diagnostics recalls

All 2,656