CLASS II Device recall · Reported 19 Oct 2016 · FDA Enforcement Report

Siemens Immulite/Immulite 2000 Systems Rubella IgM (Rum) recalled

Siemens Healthcare Diagnostics is recalling Siemens Immulite/Immulite 2000 Systems Rubella IgM (Rum), distributed nationwide in the US. The recall was initiated on 24 Aug 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesKit lots 236M, 237, 238, 239, 239L (OUS) and D236M and D238 (US)
Quantity recalledDomestic: 177 Kits; Foreign: 3435 Kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0082-2017
FDA event ID
75037
Recalling firm
Siemens Healthcare Diagnostics, Tarrytown, NY
Classification
Class II
Status
Terminated
Recall initiated
24 Aug 2016
FDA classified
13 Oct 2016
Reported
19 Oct 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

24 Aug 2016Recall initiated by company
13 Oct 2016Classified by FDA+50 days
19 Oct 2016Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Siemens received multiple customer complaints indicating an increase in false positive and indeterminate patient sample results and increased imprecision for several lots of Rubella IgM assay on the IMMMULITE 2000/IMMULITE 2000 XPi platforms.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

SIEMENS IMMULITE/IMMULITE 2000 Systems Rubella IgM (RUM), REF/Catalog Number for US: L2KRM2(D), Siemens Material Number (SMN) for US: 10374021, REF/Catalog # for OUS: L2KRM2, SMN for OUS: 10381327; Unique Device Identification Number (UDI) 00630414961996; IVD --- For in vitro diagnostic use with the IMMULITE/IMMULITE 1000/IMMULITE 2000/IMMULITE 2000 XPi Systems Analyzers-for the qualitative detection of IgM antibodies to Rubella virus in human serum or plasma (EDTA and heparinized), as an aid in the presumptive diagnosis of an acute or recent rubella infection, particularly in women of childbearing age

Where it was distributed

Worldwide Distribution n- US (Nationwide) and Foreign distribution to the following countries: Argentina, Bangladesh, Brazil, Bulgaria, China, Colombia, Costa Rica, Cyprus, Egypt, France, Germany, Greece, Guatemala, India, Indonesia, Israel, Italy, Kazakhstan, Macedonia, Mexico, Paraguay, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Africa, Syria, United Arab Emirates, and Uruguay.

Nationwide

Questions about this recall

Is Siemens Immulite/Immulite 2000 Systems Rubella IgM (Rum) recalled?

Yes. Siemens Healthcare Diagnostics recalled Siemens Immulite/Immulite 2000 Systems Rubella IgM (Rum) on 24 Aug 2016. The recall is Class II and its status is Terminated. Recall number Z-0082-2017.

Why was it recalled?

Siemens received multiple customer complaints indicating an increase in false positive and indeterminate patient sample results and increased imprecision for several lots of Rubella IgM assay on the IMMMULITE 2000/IMMULITE 2000 XPi platforms.

Which lots are affected?

Kit lots 236M, 237, 238, 239, 239L (OUS) and D236M and D238 (US)

Where was it sold?

Worldwide Distribution n- US (Nationwide) and Foreign distribution to the following countries: Argentina, Bangladesh, Brazil, Bulgaria, China, Colombia, Costa Rica, Cyprus, Egypt, France, Germany, Greece, Guatemala, India, Indonesia, Israel, Italy, Kazakhstan, Macedonia, Mexico, Paraguay, Poland, Portugal, Romania, Russian Federation, Serbia, Singapore, Slovakia, South Africa, Syria, United Arab Emirates, and Uruguay.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.