CLASS II ONGOING Device recall · Reported 1 Feb 2023 · FDA Enforcement Report

Single-Use Packs recalled by Johnson & Johnson Surgical

Johnson & Johnson Surgical Vision is recalling Single-Use Packs, distributed in 8 states. The recall was initiated on 7 Dec 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI/Lot: 05050474573352/60326033
Quantity recalled384

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1007-2023
FDA event ID
91290
Recalling firm
Johnson & Johnson Surgical Vision, Irvine, CA
Classification
Class II
Status
Ongoing
Recall initiated
7 Dec 2022
FDA classified
20 Jan 2023
Reported
1 Feb 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Dec 2022Recall initiated by company
20 Jan 2023Classified by FDA+44 days
1 Feb 2023Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Single-Use Packs, REF: OPO80

Where it was distributed

US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom

FloridaIllinoisMinnesotaMississippiNew YorkOhioTennesseeTexas

Questions about this recall

Why was it recalled?

Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Which lots are affected?

UDI-DI/Lot: 05050474573352/60326033

Where was it sold?

US: FL, MS, NY, TN, TX, IL, MN, OH. OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 91290

This product is part of a recall event; the main page for the event is Fusion Dual Pump Packs recalled by Johnson & Johnson Surgical.

More recalls from Johnson & Johnson Surgical Vision

All 18

Other Other Medical Devices recalls

All 13,884