CLASS II ONGOING Device recall · Reported 18 Jun 2025 · FDA Enforcement Report
SmartPath to dStream for 1.5T¿; Product Number recalled
Philips North America is recalling SmartPath to dStream for 1.5T¿; Product Number: 782146, distributed nationwide in the US. The recall was initiated on 29 May 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1936-2025
- FDA event ID
- 96999
- Recalling firm
- Philips North America, Cambridge, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 29 May 2025
- FDA classified
- 9 Jun 2025
- Reported
- 18 Jun 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 29 May 2025 | Recall initiated by company | |
| 9 Jun 2025 | Classified by FDA | +11 days |
| 18 Jun 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
SmartPath to dStream for 1.5T¿; Product Number: 782146;
Where it was distributed
Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.
Nationwide ArizonaCaliforniaDelawareMassachusettsMichiganNebraskaNew YorkSouth Dakota
Questions about this recall
Why was it recalled?
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or even death. This issue could also lead to property damage.
Which lots are affected?
Product Number: 782146; UDI-DI: no UDI; Serial Number: 13013;
Where was it sold?
Worldwide - US Nationwide distribution in the states of AZ, CA, DE, MA, MI, MY, NE, NY, SD and the countries of Australia, Brazil, Ecuador, Germany, Iran, Italy, Japan, Lebanon, Libya, Mexico, Poland, Portugal, Spain.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
8 other products · FDA event 96999This product is part of a recall event; the main page for the event is Intera 1.5T Achieva Nova-Dual Product Number recalled.
More recalls from Philips North America
All 531| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips Brilliance recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips IQon Spectral CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Ingenuity CT recalled by Philips North America | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Philips Brilliance CT 64: Brilliance CT 64 Channel recalled | Philips North America | Device Malfunction | MA |
| Device | CLASS II | Cardiac Workstation recalled by Philips North America | Philips North America | Labeling Errors | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |