CLASS II Device recall · Reported 6 Jan 2021 · FDA Enforcement Report

Smoothbore breathing system with exhalation port Device recalled

Intersurgical is recalling Smoothbore breathing system with exhalation port Device. The recall was initiated on 11 Nov 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: B0062001 Lot Numbers: 4200283 and 4200286 UDI Codes: (01)00081647202239(10)4200286 (01)00081647202239(10)4200283
Quantity recalled200 devices

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0753-2021
FDA event ID
86800
Recalling firm
Intersurgical, East Syracuse, NY
Classification
Class II
Status
Terminated
Recall initiated
11 Nov 2020
FDA classified
30 Dec 2020
Reported
6 Jan 2021
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Packaging Defects

Timeline

11 Nov 2020Recall initiated by company
30 Dec 2020Classified by FDA+49 days
6 Jan 2021Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The device was assembled with the incorrect exhalation port which can cause gas leakage.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Smoothbore breathing system with exhalation port 22 mm Device REF #: B0062001 Custom Anasthesia Breathing Systems; Multipac Anasthesia Breathing Systems - Product Usage: intended to administer medical gases to a patient during anesthesia. It provides both an inhalation and exhalation route and may include a connector, adaptor, and Y-piece.

Where it was distributed

All of the devices were shipped to the same distributor (McKesson).

Questions about this recall

Why was it recalled?

The device was assembled with the incorrect exhalation port which can cause gas leakage.

Which lots are affected?

Model Number: B0062001 Lot Numbers: 4200283 and 4200286 UDI Codes: (01)00081647202239(10)4200286 (01)00081647202239(10)4200283

Where was it sold?

All of the devices were shipped to the same distributor (McKesson).

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Intersurgical

All 13
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIOne-Piece Guedel Airway, size 00, ISO 5.0, Blue recalledIntersurgicalOtherNationwide
DeviceCLASS IIOne-Piece Guedel Airway, size 000, ISO 3.5, Light Green recalledIntersurgicalOtherNationwide
DeviceCLASS IIOne-piece Guedel airway, size 3, ISO 9.0, yellow recalledIntersurgicalOther10 states
DeviceCLASS IIOne-piece Guedel airway, size 2, ISO 8.0, green recalledIntersurgicalOther10 states
DeviceCLASS Ii-view video laryngoscope recalled by IntersurgicalIntersurgicalOther17 states

Other Other Medical Devices recalls

All 13,929