CLASS II Device recall · Reported 4 Apr 2018 · FDA Enforcement Report

Sofia 2 Analyzer recalled by Quidel

Quidel is recalling Sofia 2 Analyzer. The recall was initiated on 15 Dec 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Numbers: 29003956, 29004025, 29004086, 29004072, 29004188, 29004150, 29004229, 29004127, 29004232, 29004204, 29004256, 29004362, 29004155, 29004290, 29004329, 29004347, 29004350, 29004353, 29004370, 29004390, 29004396, 29004415, 29003381, 29003814, 29004183, 29004259, 29004263, 29004281, 29004317, 29004343, 29004205, 29004209, 29004001, 29004062, 29004077, 29003915, 29004036, 29004057, 29003979, 29004085, 29004144, 29004267, 29004322, 29003960, 29003991, 29004163, 29004190, 29004330, 29003992, 29004033, 29003987, 29004306, 29004325, 29004332, 29004066, 29004335, 29004373, 29004430, 29004223, 29004299, 29003999, 29004026, 29004069, 29003811, 29004164, 29004189, 29004226, 29003923, 29004160, 29004206, 29004181, 29004194, 29003584, 29004172, 29004014, 29004046, 29004352, 29004354, 29004376, 29004285, 29004377, 29004016, 29004047, 29004054, 29004196, 29003970, 29004173, 29004372, 29004252, 29004287, 29004311, 29004159, 29004356, 29004381, 29004089, 29004092, 29004270, 29004331, 29004357, 29004413, 29003578, 29004250, 29003967, 29004235, 29004319, 29003928, 29004096, 29004262, 29004280, 29004273, 29004302, 29004284, 29004278, 29004241, 29004305, 29004067, 29004125, 29004228, 29004265, 29004097, 29004303, 29004342, 29004349, 29004286, 29004098, 29004108, 29004128, 29004130, 29003947, 29004037, 29003961, 29004408, 29003636, 29003990, 29004240, 29004215, 29003738, 29004093, 29004070, 29003635, 29004006, 29004011, 29004015, 29004061, 29004073, 29004088, 29004106, 29004142, 29004148, 29004161, 29004162, 29004351, 29004391, 29004418, 29004146, 29004170, 29003978, 29004123, 29004208, 29004249, 29004315, 29004407, 29004111, 29004116, 29004243, 29004469, 29004511, 29004535, 29004548, 29003977, 29004187, 29004018, 29004039, 29004052, 29004009, 29003952, 29002222, 29004230, 29003957, 29003965, 29004359, 29004346, 29004361, 29004369, 29004378, 29004387, 29004411, 29004393, 29004441, 29004443, 29003769, 29003840, 29004147, 29004156, 29004176, 29004184, 29004186, 29004192, 29004202, 29004203, 29004210, 29004214, 29004216, 29004219, 29004222, 29004225, 29004227, 29003980, 29004048, 29004064, 29004304, 29004423, 29004283, 29004295, 29004297, 29004298, 29004310, 29004392, 29004224, 29004231, 29004558, 29004076, 29004165, 29004414, 29004274, 29004292, 29004300, 29004314, 29004318, 29004041, 29004101, 29004334, 29004380, 29004074, 29003886, 29004023, 29004081, 29004402, 29003892, 29004211, 29003936, 29004003, 29004404, 29003703, 29004153, 29004191, 29004213, 29004100, 29004040, 29004333, 29004336, 29004367, 29004291, 29004338, 29004288, 29004365, 29004239, 29004337, 29004486, 29004504, 29004508, 29004519, 29004524, 29003891, 29002538, 29004242, 29004174, 29003971, 29004012, 29003617, 29003975, 29003973, 29004145, 29003700, 29004133, 29004275, 29004326, 29004339, 29004307, 29004233, 29004257, 29004282, 29004296, 29004309, 29004323, 29004084, 29004234, 29004238, 29004254, 29004260, 29004261, 29004316, 29004447, 29004503, 29004533, 29004545, 29004552, 29004143, 29004197, 29004151, 29002356, 29002457, 29003946, 29003948, 29004017, 29004045, 29004095, 29004102, 29004104, 29004107, 29004110, 29004134, 29004137, 29004152, 29004157, 29004022, 29004382, 29004434, 29004027, 29003989, 29003995, 29004035, 29004087, 29004090, 29003993, 29004105, 29004114, 29004279, 29003688, 29004131, 29004139, 29004374, 29004360, 29004368, 29004375, 29004371, 29003731, 29004397, 29004328, 29004388, 29004010, 29004177, 29004348, 29004355, 29004366, 29004383, 29004399, 29004422, 29004425, 29004432, 29004440, 29004294, 29002598, 29003969, 29004112, 29004117, 29004412, 29004416, 29004421, 29003893, 29003895, 29003997, 29004030, 29004060, 29001731, 29004327, 29004345, 29004398, 29004405, 29004410, 29004419, 29004427, 29004439, 29004444, 29004384, 29004236, 29004272, 29004293, 29004308, 29004321, 29004251, 29004289, 29004080, 29004136, 29004271, 29004386, 29003389, 29004042, 29004078, 29003981, 29004083, 29004004, 29004051, 29004024, 29004068, 29004258, 29004409, 29004217, 29004417, 29004389, 29003313, 29003530, 29004198, 29004385, 29004029, 29004247, 29004266, 29004220, 29003954, 29004091, 29004124, 29004140, 29004401, 29004406, 29004438, 29004126, 29004400, 29004320, 29003988, 29004075, 29003976, 29004058, 29004113, 29004121, 29004158, 29003983, 29004094, 29004049, 29004071, 29003984, 29004005, 29004028, 29004044, 29004055, 29004065, 29004154, 29004180, 29004195, 29004420, 29004269, 29004344, 29004268, 29003607, 29004379, 29004403, 29004031, 29004246, 29004312, 29004301, 29004313, 29004255, 29004277, 29004264, 29004244, 29004245, 29002389, 29003275, 29003675, 29004115, 29004175, 29004179, 29004200, 29004207, 29004218, 29003996, 29004020, 29004019, 29003931, 29003998, 29003949, 29003966, 29004008, 29003994, 29003968, 29002280, 29003985, 29004043, 29003896, 29003986, 29004120, 29004433, 29004358
Quantity recalled482 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1183-2018
FDA event ID
79271
Recalling firm
Quidel, San Diego, CA
Classification
Class II
Status
Terminated
Recall initiated
15 Dec 2017
FDA classified
26 Mar 2018
Reported
4 Apr 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

15 Dec 2017Recall initiated by company
26 Mar 2018Classified by FDA+101 days
4 Apr 2018Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.

Where it was distributed

US Distribution

Questions about this recall

Why was it recalled?

The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.

Which lots are affected?

Serial Numbers: 29003956, 29004025, 29004086, 29004072, 29004188, 29004150, 29004229, 29004127, 29004232, 29004204, 29004256, 29004362, 29004155, 29004290, 29004329, 29004347, 29004350, 29004353, 29004370, 29004390, 29004396, 29004415, 29003381, 29003814, 29004183, 29004259, 29004263, 29004281, 29004317, 29004343, 29004205, 29004209, 29004001, 29004062, 29004077, 29003915, 29004036, 29004057, 29003979, 29004085, 29004144, 29004267, 29004322, 29003960, 29003991, 29004163, 29004190, 29004330, 29003992, 29004033, 29003987, 29004306, 29004325, 29004332, 29004066, 29004335, 29004373, 29004430, 29004223, 29004299, 29003999, 29004026, 29004069, 29003811, 29004164, 29004189, 29004226, 29003923, 29004160, 29004206, 29004181, 29004194, 29003584, 29004172, 29004014, 29004046, 29004352, 29004354, 29004376, 29004285, 29004377, 29004016, 29004047, 29004054, 29004196, 29003970, 29004173, 29004372, 29004252, 29004287, 29004311, 29004159, 29004356, 29004381, 29004089, 29004092, 29004270, 29004331, 29004357, 29004413, 29003578, 29004250, 29003967, 29004235, 29004319, 29003928, 29004096, 29004262, 29004280, 29004273, 29004302, 29004284, 29004278, 29004241, 29004305, 29004067, 29004125, 29004228, 29004265, 29004097, 29004303, 29004342, 29004349, 29004286, 29004098, 29004108, 29004128, 29004130, 29003947, 29004037, 29003961, 29004408, 29003636, 29003990, 29004240, 29004215, 29003738, 29004093, 29004070, 29003635, 29004006, 29004011, 29004015, 29004061, 29004073, 29004088, 29004106, 29004142, 29004148, 29004161, 29004162, 29004351, 29004391, 29004418, 29004146, 29004170, 29003978, 29004123, 29004208, 29004249, 29004315, 29004407, 29004111, 29004116, 29004243, 29004469, 29004511, 29004535, 29004548, 29003977, 29004187, 29004018, 29004039, 29004052, 29004009, 29003952, 29002222, 29004230, 29003957, 29003965, 29004359, 29004346, 29004361, 29004369, 29004378, 29004387, 29004411, 29004393, 29004441, 29004443, 29003769, 29003840, 29004147, 29004156, 29004176, 29004184, 29004186, 29004192, 29004202, 29004203, 29004210, 29004214, 29004216, 29004219, 29004222, 29004225, 29004227, 29003980, 29004048, 29004064, 29004304, 29004423, 29004283, 29004295, 29004297, 29004298, 29004310, 29004392, 29004224, 29004231, 29004558, 29004076, 29004165, 29004414, 29004274, 29004292, 29004300, 29004314, 29004318, 29004041, 29004101, 29004334, 29004380, 29004074, 29003886, 29004023, 29004081, 29004402, 29003892, 29004211, 29003936, 29004003, 29004404, 29003703, 29004153, 29004191, 29004213, 29004100, 29004040, 29004333, 29004336, 29004367, 29004291, 29004338, 29004288, 29004365, 29004239, 29004337, 29004486, 29004504, 29004508, 29004519, 29004524, 29003891, 29002538, 29004242, 29004174, 29003971, 29004012, 29003617, 29003975, 29003973, 29004145, 29003700, 29004133, 29004275, 29004326, 29004339, 29004307, 29004233, 29004257, 29004282, 29004296, 29004309, 29004323, 29004084, 29004234, 29004238, 29004254, 29004260, 29004261, 29004316, 29004447, 29004503, 29004533, 29004545, 29004552, 29004143, 29004197, 29004151, 29002356, 29002457, 29003946, 29003948, 29004017, 29004045, 29004095, 29004102, 29004104, 29004107, 29004110, 29004134, 29004137, 29004152, 29004157, 29004022, 29004382, 29004434, 29004027, 29003989, 29003995, 29004035, 29004087, 29004090, 29003993, 29004105, 29004114, 29004279, 29003688, 29004131, 29004139, 29004374, 29004360, 29004368, 29004375, 29004371, 29003731, 29004397, 29004328, 29004388, 29004010, 29004177, 29004348, 29004355, 29004366, 29004383, 29004399, 29004422, 29004425, 29004432, 29004440, 29004294, 29002598, 29003969, 29004112, 29004117, 29004412, 29004416, 29004421, 29003893, 29003895, 29003997, 29004030, 29004060, 29001731, 29004327, 29004345, 29004398, 29004405, 29004410, 29004419, 29004427, 29004439, 29004444, 29004384, 29004236, 29004272, 29004293, 29004308, 29004321, 29004251, 29004289, 29004080, 29004136, 29004271, 29004386, 29003389, 29004042, 29004078, 29003981, 29004083, 29004004, 29004051, 29004024, 29004068, 29004258, 29004409, 29004217, 29004417, 29004389, 29003313, 29003530, 29004198, 29004385, 29004029, 29004247, 29004266, 29004220, 29003954, 29004091, 29004124, 29004140, 29004401, 29004406, 29004438, 29004126, 29004400, 29004320, 29003988, 29004075, 29003976, 29004058, 29004113, 29004121, 29004158, 29003983, 29004094, 29004049, 29004071, 29003984, 29004005, 29004028, 29004044, 29004055, 29004065, 29004154, 29004180, 29004195, 29004420, 29004269, 29004344, 29004268, 29003607, 29004379, 29004403, 29004031, 29004246, 29004312, 29004301, 29004313, 29004255, 29004277, 29004264, 29004244, 29004245, 29002389, 29003275, 29003675, 29004115, 29004175, 29004179, 29004200, 29004207, 29004218, 29003996, 29004020, 29004019, 29003931, 29003998, 29003949, 29003966, 29004008, 29003994, 29003968, 29002280, 29003985, 29004043, 29003896, 29003986, 29004120, 29004433, 29004358

Where was it sold?

US Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Quidel

All 14

Other Test Kits & Diagnostics recalls

All 2,661