CLASS II Device recall · Reported 29 Jul 2020 · FDA Enforcement Report

Software 1898072 IPC upgrade v 2.7.3.0 - Product Usage: The IPC recalled

Medtronic Xomed is recalling Software 1898072 IPC upgrade v 2.7.3.0 - Product Usage: The IPC, distributed nationwide in the US. The recall was initiated on 12 Jun 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel: 1898072 Affected Lot Numbers: 0217355465, 0218377490, 0218687535, 0218928638, 0219310085,0219639840, 0217856852, 0218472653, 0218714701, 0218958034, 0219443322,0219639842, 0217858947,0218472654, 0218721203, 0218958035, 0219448447, 0219639844, 0217859045,0218665536, 0218745073, 0219004569,0219465587, 0218203285, 0218666231, 0218746389,0219201122,0219518520,0218263964, 0218666243, 0218864105, 0219255151, 0219550550, 0218293282, 0218687534, 0218900337,0219281168, 0219619754.
Quantity recalled673 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2673-2020
FDA event ID
85886
Recalling firm
Medtronic Xomed, Jacksonville, FL
Classification
Class II
Status
Terminated
Recall initiated
12 Jun 2020
FDA classified
21 Jul 2020
Reported
29 Jul 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

12 Jun 2020Recall initiated by company
21 Jul 2020Classified by FDA+39 days
29 Jul 2020Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Software 1898072 IPC upgrade v 2.7.3.0 - Product Usage: The IPC system is indicated for the incision/cutting removal, drilling and sawing of soft and hard tissue and bone in surgical procedures.

Where it was distributed

Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA,CO, CT, FL, GA, HI, ID,IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OK, PA, RI, SC, TN, TX, UT, VA, WV, WA, WI. The countries of Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, United Kingdom.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichigan

Questions about this recall

Why was it recalled?

During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.

Which lots are affected?

Model: 1898072 Affected Lot Numbers: 0217355465, 0218377490, 0218687535, 0218928638, 0219310085,0219639840, 0217856852, 0218472653, 0218714701, 0218958034, 0219443322,0219639842, 0217858947,0218472654, 0218721203, 0218958035, 0219448447, 0219639844, 0217859045,0218665536, 0218745073, 0219004569,0219465587, 0218203285, 0218666231, 0218746389,0219201122,0219518520,0218263964, 0218666243, 0218864105, 0219255151, 0219550550, 0218293282, 0218687534, 0218900337,0219281168, 0219619754.

Where was it sold?

Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA,CO, CT, FL, GA, HI, ID,IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OK, PA, RI, SC, TN, TX, UT, VA, WV, WA, WI. The countries of Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Medtronic Xomed

All 25

Other Surgical Instruments recalls

All 3,809