CLASS II Device recall · Reported 22 Apr 2020 · FDA Enforcement Report

STA - Deficient VIII, Product recalled by Diagnostica Stago

Diagnostica Stago is recalling STA - Deficient VIII, distributed nationwide in the US. The recall was initiated on 4 Mar 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll lots This cleaning sequence has been implemented in the following available software versions. Customers not possessing these software levels will be upgraded by Stago Field Support personnel: STA-R¿ Max (all serial numbers): Software version 4.05.04 or higher, and Methodology Update (MU) to enhance database version MS215. STA-R¿ Evolution (serial numbers higher than 8012756): Software version 3.04.09 or higher STA¿ Compact Max (all serial numbers): Software version 110.05 or higher.
Quantity recalled3,225 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1699-2020
FDA event ID
85167
Recalling firm
Diagnostica Stago, Parsippany, NJ
Classification
Class II
Status
Terminated
Recall initiated
4 Mar 2020
FDA classified
10 Apr 2020
Reported
22 Apr 2020
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Potency & Assay

Timeline

4 Mar 2020Recall initiated by company
10 Apr 2020Classified by FDA+37 days
22 Apr 2020Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in tests for the determination of factor VIII activity in plasma on STA-R and STA Compact analyzer platforms as follows: (A) STA-R Max, Product Codes: 58989, 58990 (B) STA-R Evolution, Product Codes: 58986, 58987, 58988 (C) STA Compact Max, Product Codes: 59026, 59027 (D) STA Compact, Product Codes: 58978, 58979

Where it was distributed

Nationwide

Nationwide

Questions about this recall

Why was it recalled?

A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels reported to be falsely increased

Which lots are affected?

All lots This cleaning sequence has been implemented in the following available software versions. Customers not possessing these software levels will be upgraded by Stago Field Support personnel: STA-R¿ Max (all serial numbers): Software version 4.05.04 or higher, and Methodology Update (MU) to enhance database version MS215. STA-R¿ Evolution (serial numbers higher than 8012756): Software version 3.04.09 or higher STA¿ Compact Max (all serial numbers): Software version 110.05 or higher.

Where was it sold?

Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.