CLASS II Device recall · Reported 1 Jan 2020 · FDA Enforcement Report

Sta Unicalibrator recalled by Diagnostica Stago

Diagnostica Stago is recalling Sta Unicalibrator (, distributed in 36 states. The recall was initiated on 18 Nov 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF: 00675 Lot Numbers: 253180, 253964, 254721, 255108 UDIs: (01)13607450006756(11)171231(17)191231(10)253180(241)00675 (01)13607450006756(11)180630(17)200630(10)253964(241)00675 (01)13607450006756(11)181130(17)201130(10)254721(241)00675 (01)13607450006756(11)190228(17)210228(10)255108(241)00675
Quantity recalled5,498

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0737-2020
FDA event ID
84402
Recalling firm
Diagnostica Stago, Parsippany, NJ
Classification
Class II
Status
Terminated
Recall initiated
18 Nov 2019
FDA classified
23 Dec 2019
Reported
1 Jan 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

18 Nov 2019Recall initiated by company
23 Dec 2019Classified by FDA+35 days
1 Jan 2020Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.

Product as listed by the FDA

STA UNICALIBRATOR (ref. 00675)

Where it was distributed

US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK

AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaIowaIllinoisKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouri

Questions about this recall

Why was it recalled?

A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.

Which lots are affected?

REF: 00675 Lot Numbers: 253180, 253964, 254721, 255108 UDIs: (01)13607450006756(11)171231(17)191231(10)253180(241)00675 (01)13607450006756(11)180630(17)200630(10)253964(241)00675 (01)13607450006756(11)181130(17)201130(10)254721(241)00675 (01)13607450006756(11)190228(17)210228(10)255108(241)00675

Where was it sold?

US: MA ME VT CT NJ NY PA DC MD VA NC SC GA FL AL TN MS KY OH IN MI IA WI MN SD ND MT IL MO KS LA AR OK TX CO AZ NV CA HI OR WA AK

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 84402

This product is part of a recall event; the main page for the event is Sta System Conrol N + P recalled by Diagnostica Stago.

ReportedClassRecallCompanyReasonWhere
CLASS IISta System Conrol N + P recalled by Diagnostica StagoDiagnostica StagoOther36 states

More recalls from Diagnostica Stago

All 27

Other Other Medical Devices recalls

All 13,884