CLASS II Device recall · Reported 5 Aug 2020 · FDA Enforcement Report

StealthStation" S7 System with Polaris Spectra Camera recalled

Medtronic Navigation is recalling StealthStation" S7 System with Polaris Spectra Camera, distributed nationwide in the US. The recall was initiated on 5 Jun 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesStealthStation S7 Serial Number N06164844 with Polaris Spectra Camera Serial Number P7-01896; and StealthStation S7 Serial Number N07308697 with Polaris Spectra Camera Serial Number P7-18852.
Quantity recalled2

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2707-2020
FDA event ID
85901
Recalling firm
Medtronic Navigation, Louisville, CO
Classification
Class II
Status
Terminated
Recall initiated
5 Jun 2020
FDA classified
27 Jul 2020
Reported
5 Aug 2020
Type
Voluntary: Firm initiated
Customers notified by
Telephone
Reason category
Labeling Errors

Timeline

5 Jun 2020Recall initiated by company
27 Jul 2020Classified by FDA+52 days
5 Aug 2020Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.

Product as listed by the FDA

StealthStation" S7 System with Polaris Spectra Camera part number PSU 9733437 - Product Usage: are intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.

Where it was distributed

US Nationwide distribution including in the states of Alabama and Massachusetts.

Nationwide AlabamaMassachusetts

Questions about this recall

Why was it recalled?

There is a potential for specific camera units to be affected by a characterization issue during manufacturing that results in inaccurate navigation while utilizing both the active and passive trackers simultaneously. While no injuries or illnesses have been reported, If impacted cameras were to be used with both active and passive trackers simultaneously, it would result in system navigation inaccuracy of 6mm.

Which lots are affected?

StealthStation S7 Serial Number N06164844 with Polaris Spectra Camera Serial Number P7-01896; and StealthStation S7 Serial Number N07308697 with Polaris Spectra Camera Serial Number P7-18852.

Where was it sold?

US Nationwide distribution including in the states of Alabama and Massachusetts.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.