CLASS II ONGOING Device recall · Reported 31 Dec 2025 · FDA Enforcement Report
The Stryker CranialMask Tracker is a single-use device recalled
Howmedica Osteonics is recalling The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed, distributed nationwide in the US. The recall was initiated on 12 Nov 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0963-2026
- FDA event ID
- 97988
- Recalling firm
- Howmedica Osteonics, Mahwah, NJ
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 12 Nov 2025
- FDA classified
- 19 Dec 2025
- Reported
- 31 Dec 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Masks, Gloves & PPE
Timeline
| 12 Nov 2025 | Recall initiated by company | |
| 19 Dec 2025 | Classified by FDA | +37 days |
| 31 Dec 2025 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.
Questions about this recall
Why was it recalled?
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
Which lots are affected?
Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;
Where was it sold?
Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Howmedica Osteonics
All 244| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Axle Pin Assy Large Elbow Cobalt Chrome recalled by Howmedica Osteonics | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Triathlon Hinge Insert Size 4 5612-p-411 Triathlon Hinge recalled | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Scorpio Recessed Metal Backed Patella 3045 recalled | Howmedica Osteonics | Foreign Material | Nationwide |
| Device | CLASS II | Bearing And Axle Bushing Assy Std Elbow C-m113-1-301 Krh recalled | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Scorpio + Mobile Bearing Pcs 78-5-0910 System One Bipolar recalled | Howmedica Osteonics | Other | Nationwide |
Other Masks, Gloves & PPE recalls
All 450| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Opportunity Pro recalled by American Nitrile Operations | American Nitrile Operations | Device Malfunction | Nationwide |
| Device | CLASS II | Opportunity Flex recalled by American Nitrile Operations | American Nitrile Operations | Device Malfunction | Nationwide |
| Device | CLASS II | Performance Pro recalled by American Nitrile Operations | American Nitrile Operations | Device Malfunction | Nationwide |
| Device | CLASS II | P28 Phanton Intramedullary Nail System recalled by Paragon 28 | Paragon 28 | Sterility | Nationwide |
| Device | CLASS II | Cyberknife Treatment Deliver System equipped recalled | Accuray | Device Malfunction | Nationwide |