CLASS II ONGOING Device recall · Reported 31 Dec 2025 · FDA Enforcement Report

The Stryker CranialMask Tracker is a single-use device recalled

Howmedica Osteonics is recalling The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed, distributed nationwide in the US. The recall was initiated on 12 Nov 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;
Quantity recalled320 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0963-2026
FDA event ID
97988
Recalling firm
Howmedica Osteonics, Mahwah, NJ
Classification
Class II
Status
Ongoing
Recall initiated
12 Nov 2025
FDA classified
19 Dec 2025
Reported
31 Dec 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

12 Nov 2025Recall initiated by company
19 Dec 2025Classified by FDA+37 days
31 Dec 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.

Nationwide

Questions about this recall

Why was it recalled?

When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

Which lots are affected?

Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.