CLASS II ONGOING Device recall · Reported 21 Oct 2026 · FDA Enforcement Report
TS Universal Baseplate Finishing Tray, Triathlon recalled
Howmedica Osteonics is recalling TS Universal Baseplate Finishing Tray, Triathlon. The recall was initiated on 16 Sep 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0164-2027
- FDA event ID
- 99883
- Recalling firm
- Howmedica Osteonics, Mahwah, NJ
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 16 Sep 2026
- FDA classified
- 15 Oct 2026
- Reported
- 21 Oct 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Other Medical Devices
Timeline
| 16 Sep 2026 | Recall initiated by company | |
| 15 Oct 2026 | Classified by FDA | +29 days |
| 21 Oct 2026 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.
Where it was distributed
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Questions about this recall
Why was it recalled?
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Which lots are affected?
Catalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Where was it sold?
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 27 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
3 other products · FDA event 99883More recalls from Howmedica Osteonics
All 248| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TS Misc. Instruments Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | TS Femoral Prep Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | Triathlon Hinge Femoral Prep Tray recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | Axle Pin Assy Large Elbow Cobalt Chrome recalled by Howmedica Osteonics | Howmedica Osteonics | Other | Nationwide |
| Device | CLASS II | Triathlon Hinge Insert Size 4 5612-p-411 Triathlon Hinge recalled | Howmedica Osteonics | Other | Nationwide |
Other Other Medical Devices recalls
All 13,929| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Rapicide PA Part A High-Level Disinfectant recalled by Steris | Steris | Other | Nationwide |
| Device | CLASS II | NeuroStar Advanced Therapy System recalled by Neuronetics | Neuronetics | Other | Nationwide |
| Device | CLASS II | TS Misc. Instruments Tray, Triathlon recalled by Howmedica Osteonics | Howmedica Osteonics | Sterility | NJ |
| Device | CLASS II | BD Phoenix Panel Nmic recalled by Becton Dickinson | Becton Dickinson | Device Malfunction | Nationwide |
| Device | CLASS II | BD Phoenix Panel Nmic recalled by Becton Dickinson | Becton Dickinson | Device Malfunction | Nationwide |