CLASS II Device recall · Reported 15 May 2013 · FDA Enforcement Report

Stryker Irrigator Reusable Tips: Regular Tip with 2 Hole recalled

Stryker Endoscopy is recalling Stryker Irrigator Reusable Tips: Regular Tip with 2 Hole, Metal Finish, distributed nationwide in the US. The recall was initiated on 27 Mar 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesManufacturer Part Number: 250-070-472
Quantity recalled147,921 - all sizes

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1227-2013
FDA event ID
64682
Recalling firm
Stryker Endoscopy, San Jose, CA
Classification
Class II
Status
Terminated
Recall initiated
27 Mar 2013
FDA classified
3 May 2013
Reported
15 May 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Foreign Material

Timeline

27 Mar 2013Recall initiated by company
3 May 2013Classified by FDA+37 days
15 May 2013Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

Stryker Irrigator Reusable Tips: 2mm X 32cm Regular Tip with 2 Hole, Metal Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.

Where it was distributed

Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.

Nationwide

Questions about this recall

Is Stryker Irrigator Reusable Tips: Regular Tip with 2 Hole, Metal Finish recalled?

Yes. Stryker Endoscopy recalled Stryker Irrigator Reusable Tips: Regular Tip with 2 Hole, Metal Finish on 27 Mar 2013. The recall is Class II and its status is Terminated. Recall number Z-1227-2013.

Why was it recalled?

The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C.

Which lots are affected?

Manufacturer Part Number: 250-070-472

Where was it sold?

Worldwide Distribution: USA (nationwide) and countries of: Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, Phillipines, Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brasil, Romania, South Africa, Chile, Korea, China.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Endoscopy

All 238

Other Surgical Instruments recalls

All 3,809