CLASS II Device recall · Reported 4 Apr 2018 · FDA Enforcement Report

The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart recalled

Stryker Instruments Div. of Stryker is recalling The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control, distributed nationwide in the US. The recall was initiated on 16 Oct 2017 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Number(s) Affected: 10897, 10898, 10899, 10900, 10901, 10902, 10903, 10904, 10905, 10906, 10907, 10908, 10909, 10910, 10911, 10912, 10913, 10914, 10917, 10919, 10920, 10921, 10922, 10923, 10925, 10926, 10927, 10928. Product Code (UDI): (01)04546540503824
Quantity recalled28

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1216-2018
FDA event ID
79335
Recalling firm
Stryker Instruments Div. of Stryker, Portage, MI
Classification
Class II
Status
Terminated
Recall initiated
16 Oct 2017
FDA classified
27 Mar 2018
Reported
4 Apr 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay
Device type
Spinal Implants

Timeline

16 Oct 2017Recall initiated by company
27 Mar 2018Classified by FDA+162 days
4 Apr 2018Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ Navigation System because the settings on the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a pathway into the pedicle. The instrument has three types of buttons to remotely control the software: the UP, DOWN and SELECT buttons. The Pedicle Feeler is laser marked in 10 mm intervals to help indicate the depth in which the feeler tip has been inserted into the pedicle. Next to the navigational information, these depth indicators on the feeler tip are also helpful in determining the appropriate pedicle screw length.

Where it was distributed

International , US and US Govt.

Nationwide

Questions about this recall

Is The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control recalled?

Yes. Stryker Instruments Div. of Stryker recalled The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control on 16 Oct 2017. The recall is Class II and its status is Terminated. Recall number Z-1216-2018.

Why was it recalled?

Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ Navigation System because the settings on the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break.

Which lots are affected?

Serial Number(s) Affected: 10897, 10898, 10899, 10900, 10901, 10902, 10903, 10904, 10905, 10906, 10907, 10908, 10909, 10910, 10911, 10912, 10913, 10914, 10917, 10919, 10920, 10921, 10922, 10923, 10925, 10926, 10927, 10928. Product Code (UDI): (01)04546540503824

Where was it sold?

International , US and US Govt.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Instruments Div. of Stryker

All 161

Other Spinal Implants recalls

All 1,157