CLASS II Device recall · Reported 27 Jul 2022 · FDA Enforcement Report

Triton Canister System (finished part numbers FG recalled

Stryker Instruments Div. of Stryker is recalling Triton Canister System (finished part numbers FG 12009, for canisters, and FG 12019 for, distributed nationwide in the US. The recall was initiated on 22 Jan 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI GTIN: 10859506006101, 10859506006125 All Canister Scanning Labels with a date of manufacture prior to to 2019-12-24
Quantity recalled16,700 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1402-2022
FDA event ID
90464
Recalling firm
Stryker Instruments Div. of Stryker, Portage, MI
Classification
Class II
Status
Terminated
Recall initiated
22 Jan 2020
FDA classified
19 Jul 2022
Reported
27 Jul 2022
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction
Device type
Medical Software

Timeline

22 Jan 2020Recall initiated by company
19 Jul 2022Classified by FDA+909 days
27 Jul 2022Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.

Where it was distributed

US Nationwide

Nationwide

Questions about this recall

Why was it recalled?

Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.

Which lots are affected?

UDI-DI GTIN: 10859506006101, 10859506006125 All Canister Scanning Labels with a date of manufacture prior to to 2019-12-24

Where was it sold?

US Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Instruments Div. of Stryker

All 161

Other Medical Software recalls

All 908