CLASS II Device recall · Reported 3 Jul 2013 · FDA Enforcement Report
SV Elastomeric Infusion Device recalled by Baxter Healthcare
Baxter Healthcare is recalling SV Elastomeric Infusion Device, distributed nationwide in the US. The recall was initiated on 7 Jun 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1605-2013
- FDA event ID
- 65365
- Recalling firm
- Baxter Healthcare, Round Lake, IL
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 7 Jun 2013
- FDA classified
- 27 Jun 2013
- Reported
- 3 Jul 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 7 Jun 2013 | Recall initiated by company | |
| 27 Jun 2013 | Classified by FDA | +20 days |
| 3 Jul 2013 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality. Complaint rates for ruptures have since decreased. Baxter wants to ensure that no affected product remains in the field.
Product as listed by the FDA
Brand Name: SV Elastomeric Infusion Device. Indicated for the intravenous administration of medications.
Where it was distributed
Distributed Nationwide and in Puerto Rico.
Questions about this recall
Is SV Elastomeric Infusion Device recalled?
Yes. Baxter Healthcare recalled SV Elastomeric Infusion Device on 7 Jun 2013. The recall is Class II and its status is Terminated. Recall number Z-1605-2013.
Why was it recalled?
Baxter Healthcare Corporation (Baxter) is issuing an Urgent Product Recall for the products listed above due to an increase in rupture complaints. Baxter has implemented product and process improvements which have substantially improved product quality. Complaint rates for ruptures have since decreased. Baxter wants to ensure that no affected product remains in the field.
Which lots are affected?
Product Codes: 2C1071KJP, 2C1073KJP, 2C1075KJP Lot Codes: 08J078 08K002 08K005 08K007 08K054 08K066 08M001 08M025 08M028 08M039 08N003 08N005 08N032 09A004 09A040 09A042 09A064 98025 98027 98028 98070 09(018 09C023 09(024 09C045 09C047 09(075 90029 90031 90033 90065 90069 90086 09E026 09E027 09E031 09F023 09F024 09F025 09F040 09F041 09F042 09G013 09G061 09G062 09H004 09H008 09H035 09H036 09H070 09H082 09J006 09J007 09J074 09J075 09K017 09K018 09K021 09K022 09K086 09K087 09M002 09M005 09M037 09M038 09M039 09M078 09N027 09N028 09N029 09N034 10A019 10A020 10A021 10A044 108010 108029 10C002 10C003 10C006 10C078 10C079 10C099 10C103 10C111 100026 100027 100028 100058 100062 100063 10E003 10E008 10E009 10E067 10G005 10G006 10G007 10G008 10G078 10G079 10H014 10H019 10H020 10H022 10H054 10H056 10H057 10H060 10H069 10H071 10H105 10H107 10J022 10J023 10J026 10J028 10J029 10J070 10J071 10J074 10K043 10K044 10K047 10K049 10K051 10K061 10K065 10K110 10M056 10M058 10M059 10M060 10N006 10N024 10N028 10N078 11A022 11A024 11A030 11A052 11A074 11A075 08K001 08K003 08K053 08K077 08M002 08N007 08N029 08N030 08N034 08N069 08N070 09A001 09A063 98026 98069 98071 09C046 90025 90032 90070 09E030 09F022 09F026 09G010 09H005 09H037 09J005 09J042 09J043 09J044 09J045 09J089 09K020 09K050 09M001 09M041 09N032 09N041 10A017 10A022 10A046 108014 108024 10C076 10C100 10C102 100025 100029 10E002 10E007 10E045 10E055 10G001 10G003 10G004 10H024 10H058 10H059 10H072 10J021 10J024 10J027 10J072 10K046 10K048 10K063 10K064 10M034 10M057 10N023 10N077 11A023 11A031 11A054 08K067 08M003 09A006 98078 90026 90085 09E052 09H001 09J071 09K053 09M003 09M077 108025 10C005 10C080 10C101 100059 10E053 10H023 10H106 10J030 10J045 10J069 10K066 10M086 10N027 11A025 Expiration Dates: 06/30/13, 12/31/13, 12/31/14, 06/30/14, 12/31/14, 06/30/15,12/31/15, 06/30/15
Where was it sold?
Distributed Nationwide and in Puerto Rico.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Baxter Healthcare
All 564| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Drug | CLASS I | Dextrose Injection, USP, 70 % recalled by Baxter Healthcare | Baxter Healthcare | Particulate Matter | Nationwide |
| Drug | CLASS I | 9% Sodium Chloride Injection USP, Viaflex recalled by Baxter HealthcareSodium Chloride | Baxter Healthcare | Particulate Matter | IL |
| Device | CLASS II | Baxter Novum IQ Syringe Pump, Loaner recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Baxter Novum IQ Syringe Pump recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |
| Drug | CLASS II | Clindamycin Phosphate in 9% Sodium Chloride Injection recalledClindamycin Phosphate | Baxter Healthcare | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |