CLASS II Device recall · Reported 20 Jul 2016 · FDA Enforcement Report

Target Detachable Coils: Target Helical Nano recalled

Stryker Neurovascular is recalling Target Detachable Coils: Target Helical Nano, distributed nationwide in the US. The recall was initiated on 6 Jun 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesMODEL Number:M0035452060; Lot Number/Expiration Date: 18777380 31-Dec-18 18538894 31-Aug-18 18726354 30-Nov-18 18830541 31-Jan-19 18802659 31-Jan-19 18776332 31-Dec-18 18499021 31-Jul-18 18743910 30-Nov-18 18725944 30-Nov-18 18454649 30-Jun-18 18794568 31-Jan-19 18538832 31-Aug-18 18829512 31-Jan-19 18801344 31-Jan-19 18709049 30-Nov-18 18693095 30-Nov-18 18499207 31-Jul-18 19032899 30-Apr-19 19047487 31-May-19 18852469 28-Feb-19 18954631 31-Mar-19 18906669 28-Feb-19 18877336 28-Feb-19 19037939 31-May-19 18973118 30-Apr-19 18899998 28-Feb-19 18900840 28-Feb-19 19046456 31-May-19 18972772 30-Apr-19 18955537 31-Mar-19 19005724 30-Apr-19 18935230 31-Mar-19 18925282 31-Mar-19 18876468 28-Feb-19 19032992 30-Apr-19.
Quantity recalled45,302 (8,188 in the U.S.) - total, all model numbers

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2181-2016
FDA event ID
74425
Recalling firm
Stryker Neurovascular, Fremont, CA
Classification
Class II
Status
Terminated
Recall initiated
6 Jun 2016
FDA classified
11 Jul 2016
Reported
20 Jul 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals

Timeline

6 Jun 2016Recall initiated by company
11 Jul 2016Classified by FDA+35 days
20 Jul 2016Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

Target Detachable Coils: TARGET HELICAL NANO 2MM X 6CM MODEL Number:M0035452060 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Where it was distributed

Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.

Nationwide

Stores named: Target

Questions about this recall

Is Target Detachable Coils: Target Helical Nano recalled?

Yes. Stryker Neurovascular recalled Target Detachable Coils: Target Helical Nano on 6 Jun 2016. The recall is Class II and its status is Terminated. Recall number Z-2181-2016.

Why was it recalled?

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Which lots are affected?

MODEL Number:M0035452060; Lot Number/Expiration Date: 18777380 31-Dec-18 18538894 31-Aug-18 18726354 30-Nov-18 18830541 31-Jan-19 18802659 31-Jan-19 18776332 31-Dec-18 18499021 31-Jul-18 18743910 30-Nov-18 18725944 30-Nov-18 18454649 30-Jun-18 18794568 31-Jan-19 18538832 31-Aug-18 18829512 31-Jan-19 18801344 31-Jan-19 18709049 30-Nov-18 18693095 30-Nov-18 18499207 31-Jul-18 19032899 30-Apr-19 19047487 31-May-19 18852469 28-Feb-19 18954631 31-Mar-19 18906669 28-Feb-19 18877336 28-Feb-19 19037939 31-May-19 18973118 30-Apr-19 18899998 28-Feb-19 18900840 28-Feb-19 19046456 31-May-19 18972772 30-Apr-19 18955537 31-Mar-19 19005724 30-Apr-19 18935230 31-Mar-19 18925282 31-Mar-19 18876468 28-Feb-19 19032992 30-Apr-19.

Where was it sold?

Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Neurovascular

All 109

Other Heart & Vascular Devices recalls

All 6,199