CLASS II ONGOING Device recall · Reported 2 Aug 2023 · FDA Enforcement Report

Trans-ray IAB, Intra-Aortic Balloon Catheter recalled by Datascope

Datascope is recalling Trans-ray IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter, distributed nationwide in the US. The recall was initiated on 23 Jun 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesProduct Description UDI 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES 10607567107656 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES 10607567107663 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES 10607567111711 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES 10607567111728 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs 10607567107684 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs 10607567111810
Quantity recalled321,609 total kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2218-2023
FDA event ID
92527
Recalling firm
Datascope, Fairfield, NJ
Classification
Class II
Status
Ongoing
Recall initiated
23 Jun 2023
FDA classified
21 Jul 2023
Reported
2 Aug 2023
Type
Voluntary: Firm initiated
Customers notified by
N/A
Reason category
Packaging Defects

Timeline

23 Jun 2023Recall initiated by company
21 Jul 2023Classified by FDA+28 days
2 Aug 2023Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES D684-00-0545-01 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES D684-00-0546-01 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES D684-00-0607 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES D684-00-0608 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs D884-00-0019-18 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs D884-00-0019-26

Where it was distributed

Domestic distribution nationwide. International distribution worldwide.

Nationwide

Questions about this recall

Why was it recalled?

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Which lots are affected?

Product Description UDI 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES 10607567107656 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES 10607567107663 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES 10607567111711 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES 10607567111728 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs 10607567107684 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs 10607567111810

Where was it sold?

Domestic distribution nationwide. International distribution worldwide.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

8 other products · FDA event 92527

This product is part of a recall event; the main page for the event is Linear 7..IAB, Intra-Aortic Balloon Catheter recalled by Datascope.

More recalls from Datascope

All 134
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIIBattery Charging Station recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IICS100 IABP recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IICS300 IABP recalled by DatascopeDatascopeDevice MalfunctionNationwide
DeviceCLASS IIICardiosave Hybrid recalled by DatascopeDatascopeOtherNationwide
DeviceCLASS IIICardiosave Rescue recalled by DatascopeDatascopeOtherNationwide

Other Heart & Vascular Devices recalls