CLASS II Device recall · Reported 31 Aug 2016 · FDA Enforcement Report

TrueBeam¿ Radiotherapy Delivery System recalled by Vision RT

Vision RT is recalling TrueBeam¿ Radiotherapy Delivery System and EDGE" Radiotherapy Delivery System, K140528. The recall was initiated on 30 Jun 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial Number of the device(s) distributed 248-016 248-021 248-022 248-030 248-038 248-041 248-063 248-064 249-1514 249-1516 248-071 249-0239 248-073 248-080 248-084 249-0043 249-0055 249-0056 249-0058 249-0069 249-0130 249-0137 249-1509 249-0141 249-0275 249-0152 249-0162 249-0169 249-0251 249-0253 249-0254 249-0288 249-1503 249-1512 249-1518 249-TBD 249-0061 249-0131 249-0256 249-0289 249-0290 249-TBD 248-083 249-0273 249-0276 249-1500 249-1519
Quantity recalled47

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2668-2016
FDA event ID
74607
Recalling firm
Vision RT, London
Classification
Class II
Status
Terminated
Recall initiated
30 Jun 2016
FDA classified
25 Aug 2016
Reported
31 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Other Reasons

Timeline

30 Jun 2016Recall initiated by company
25 Aug 2016Classified by FDA+56 days
31 Aug 2016Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Vision RT Ltd received three (3) reports from different sites that discovered the following issue, unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up for the Optical Surface Monitoring System [OSMS]. 1. CRM 3194, No Serious Injury, No MDR, Aware date: Dec 11, 2015 2. CRM 3321, No Patient Involved, No MDR, Aware date: Jan 19, 2016 3. CRM 4570, No Serious Injury, No MDR, Aware date: March 15, 2016 No patient harm was reported in any of these cases.

Product as listed by the FDA

TrueBeam¿ Radiotherapy Delivery System and EDGE" Radiotherapy Delivery System, K140528. The TrueBeam and Edge Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.

Where it was distributed

Product distributed in US. The foreign countries where the product was distributed are as follows: Canada, Germany, Japan, Sweden, Switzerland, Taiwan, and United Kingdom. No product was distributed to US government accounts.

Questions about this recall

Is TrueBeam¿ Radiotherapy Delivery System and EDGE" Radiotherapy Delivery System, K140528 recalled?

Yes. Vision RT recalled TrueBeam¿ Radiotherapy Delivery System and EDGE" Radiotherapy Delivery System, K140528 on 30 Jun 2016. The recall is Class II and its status is Terminated. Recall number Z-2668-2016.

Why was it recalled?

Vision RT Ltd received three (3) reports from different sites that discovered the following issue, unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up for the Optical Surface Monitoring System [OSMS]. 1. CRM 3194, No Serious Injury, No MDR, Aware date: Dec 11, 2015 2. CRM 3321, No Patient Involved, No MDR, Aware date: Jan 19, 2016 3. CRM 4570, No Serious Injury, No MDR, Aware date: March 15, 2016 No patient harm was reported in any of these cases.

Which lots are affected?

Serial Number of the device(s) distributed 248-016 248-021 248-022 248-030 248-038 248-041 248-063 248-064 249-1514 249-1516 248-071 249-0239 248-073 248-080 248-084 249-0043 249-0055 249-0056 249-0058 249-0069 249-0130 249-0137 249-1509 249-0141 249-0275 249-0152 249-0162 249-0169 249-0251 249-0253 249-0254 249-0288 249-1503 249-1512 249-1518 249-TBD 249-0061 249-0131 249-0256 249-0289 249-0290 249-TBD 248-083 249-0273 249-0276 249-1500 249-1519

Where was it sold?

Product distributed in US. The foreign countries where the product was distributed are as follows: Canada, Germany, Japan, Sweden, Switzerland, Taiwan, and United Kingdom. No product was distributed to US government accounts.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Vision RT

All 6

Other Other Medical Devices recalls

All 13,778