CLASS II Device recall · Reported 18 Mar 2020 · FDA Enforcement Report

VaccZyme Human Anti-Haemophilus influenza type b Enzyme recalled

The Binding Site Group is recalling VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro, distributed nationwide in the US. The recall was initiated on 12 Aug 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 05051700006767. Lot numbers 413170 and 435314.
Quantity recalled462 kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1469-2020
FDA event ID
84899
Recalling firm
The Binding Site Group, Birmingham
Classification
Class II
Status
Terminated
Recall initiated
12 Aug 2018
FDA classified
11 Mar 2020
Reported
18 Mar 2020
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Potency & Assay

Timeline

12 Aug 2018Recall initiated by company
11 Mar 2020Classified by FDA+577 days
18 Mar 2020Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

VaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of M.A., C.A., MO., NC, MN, VA, WA, and IL and country of Canada.

Nationwide IllinoisMinnesotaMissouriNorth CarolinaVirginiaWashington

Questions about this recall

Why was it recalled?

A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.

Which lots are affected?

UDI: 05051700006767. Lot numbers 413170 and 435314.

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of M.A., C.A., MO., NC, MN, VA, WA, and IL and country of Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from The Binding Site Group

All 29

Other Test Kits & Diagnostics recalls

All 2,671